Serology I Positive Control (Serology I Control +)
Serology I Postive Control is intended for use with in vitro assays for determination of antibodies to Human Immunodeficiency Virus Type 1 and 2 (HIV-1/2), antibodies to Human…
Serology I Postive Control is intended for use with in vitro assays for determination of antibodies to Human Immunodeficiency Virus Type 1 and 2 (HIV-1/2), antibodies to Human T-Lymphotropic Virus Type 1 and 2 (HTLV-1/2), antibodies to Hepatitis C Virus (HCV), Hepatitis B Surface Antigen (HBsAg), IgG antibodies to Hepatitis B Core Antibodies (HBc) and IgG antibodies to Treponema pallidum. Serology I Positive Control is unassayed without target values and is suitable for use on many analysers. These controls must not be used as a substitute for the mandatory manufacturer’s kit controls provided with the assay. Serology I Positive Control is helpful in determining the precision of testing systems and in identifying sources of variation.
Who Else Makes It
Concentrated in Jilin (1), Guangdong (1).
| Manufacturer | City | Province | Registered as | Device types |
|---|---|---|---|---|
| Dirui Industrial Co., Ltd. | Changchun | Jilin | Contract manufacturer | 84 |
| Phase Scientific (Shenzhen) Company Limited | Shenzhen | Guangdong | Repacker / relabeller | 8 |
Record
| Listed by | Randox Laboratories Limited, United Kingdom |
|---|---|
| FDA product code | OHQ |
| Device type | Multi-Analyte Controls Unassayed |
| Device class | Class 1 |
| Regulation | 21 CFR 862.1660 |
| Establishment | Contract manufacturer, Specification developer, Repacker / relabeller, Manufacturer |
| Premarket notification | Exempt |
| Record published | 2020-05-11 |
| Record revised | 2020-12-21 (version 2) |
Specifications
| MRI safety | Labeling does not contain MRI Safety Information |
|---|---|
| Single use | Yes |
| Prescription | Prescription only |
| Supplied as a kit | Yes |
| Label carries | lot or batch number, expiry date |
| Storage | Storage environment temperature 2 to 8 °C |
Catalogue Numbers
| Catalogue number | Size | Status |
|---|---|---|
| SR10352 | In distribution |
Same Product Code
| Product | Sold by | Catalogue numbers |
|---|---|---|
| Exact Diagnostics | Exact Diagnostics | 64 |
| NATtrol™ | Zeptometrix LLC. | 54 |
| QC Sets and Panels: Helix Elite™ | Microbiologics, Inc. | 51 |
| Helix Elite™ Molecular Standards: Inactivated Standard | Microbiologics, Inc. | 41 |
| Multichem S Plus | Techno-Path Manufacturing Ltd. | 35 |
| Virotrol | Bio-Rad Laboratories Inc. | 27 |
| Helix Elite™ Molecular Standards: Synthetic Standard | Microbiologics, Inc. | 25 |
| Multichem IA Plus | Techno-Path Manufacturing Ltd. | 17 |
| D3 | Diagnostic Hybrids, Inc. | 14 |
| Liquichek | Bio-Rad Laboratories Inc. | 10 |
| Multichem IA | Techno-Path Manufacturing Ltd. | 8 |
| QC Sets and Panels: Kwik-Stik™ 2 Pack | Microbiologics, Inc. | 8 |
Showing 12 of 107. All products under Multi-Analyte Controls Unassayed
Recalls Naming This Product
1 recall names this product by its own identifier, the UDI on its label.
| Recalled by | Identified by | Initiated | FDA class | Root cause | Reason | Quantity | Closed |
|---|---|---|---|---|---|---|---|
| Randox Laboratories Limited | UDI-DI | 2022-08-01 | Class II | Process change control | ***Update** :HBsAg analyte is also Non-Reactive on the Ortho Vitros analyser as well as the Beckman Coulter DxI. Randox Serology I Positive Control will test negative (Non-Reactive) for Marker HBsAg resulting in delay in reporting patient | 2 units | 2024-12-13 |
Recalls
1 of these is Class III. Class III is unlikely to cause harm at all → typically a labelling or documentation fault. Volume alone is not a quality signal: a manufacturer that ships more recalls more.
| Initiated | FDA class | Root cause | Reason | Quantity |
|---|---|---|---|---|
| 2023-11-06 | Class III | Process control | A decrease in the reactivity of HSV Type 1/2 lgM in the Serology ToRCH lgM Positive Control when tested on the DiaSorin Liaison XL. | 1 unit |
Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.