Illumigene Mycoplasma
"The illumigene Mycoplasma DNA amplification assay, performed on the illumipro-10, is a qualitative in vitro diagnostic test for the direct detection of DNA from Mycoplasma pneumoniae in…
“The illumigene Mycoplasma DNA amplification assay, performed on the illumipro-10, is a qualitative in vitro diagnostic test for the direct detection of DNA from Mycoplasma pneumoniae in human throat and nasopharyngeal swabs obtained from patients suspected of having Mycoplasma pneumoniae infection. The illumigene Mycoplasma assay utilizes loop-mediated isothermal DNA amplification (LAMP) technology to detect Mycoplasma pneumoniae by targeting a segment of the Mycoplasma pneumoniae genome. Results from the illumigene Mycoplasma DNA amplification assay should be used in conjunction with clinical presentation, other laboratory findings, and epidemiological risk factors as an aid in the diagnosis of Mycoplasma infection and should not be used as the sole basis for treatment or other patient management.
Who Else Makes It
Concentrated in Jiangsu (4), Shandong (3), Zhejiang (2). 3 of them are registered as contract manufacturers, able to build to a buyer’s own specification.
| Manufacturer | City | Province | Registered as | Device types |
|---|---|---|---|---|
| Zhejiang Orient Gene Biotech | Huzhou | Zhejiang | Contract manufacturer | 86 |
| Guangzhou Four E’s Scientific Co., Ltd. | Guangzhou | Guangdong | Contract manufacturer | 61 |
| Sanmina | Kunshan | Jiangsu | Contract manufacturer | 15 |
| Biobase Biodustry(Shandong)Co., Ltd. | Jinan | Shandong | Manufacturer | 70 |
| Infitek Co., Ltd. | Shanghai | Shanghai | Manufacturer | 49 |
| Labpeak Co., Ltd. | Jinan | Shandong | Manufacturer | 43 |
| Scitek Global Co., Ltd. | Jinan | Shandong | Manufacturer | 43 |
| Bioteke Corporation (Wuxi) Co., Ltd. | Wuxi | Jiangsu | Manufacturer | 21 |
| Getein Biotech, Inc. | Nanjing | Jiangsu | Manufacturer | 19 |
| Xian Tianlong Science AND Technology Co., Ltd. | Xi’An | Shaanxi | Manufacturer | 8 |
| Hangzhou Bioer Technology Co., Ltd. | Hangzhou | Zhejiang | Manufacturer | 6 |
| Sansure Biotech Inc. | Changsha | Hunan | Manufacturer | 4 |
Showing 12 of 15. All 15 manufacturers in China registered for Real Time Nucleic Acid Amplification System
Record
| Listed by | Meridian Bioscience, Inc., United States |
|---|---|
| FDA product code | OOI |
| Device type | Real Time Nucleic Acid Amplification System |
| Device class | Class 2 |
| Regulation | 21 CFR 862.2570 |
| Establishment | Contract manufacturer, Specification developer, Repacker / relabeller, Complaint-file establishment, Manufacturer |
| 510(k) | K123423 |
| Record published | 2016-09-23 |
| Record revised | 2020-06-29 (version 4) |
Specifications
| MRI safety | Labeling does not contain MRI Safety Information |
|---|---|
| Single use | Yes |
| Prescription | Prescription only |
| Supplied as a kit | Yes |
| Label carries | lot or batch number, expiry date |
| Storage | Handling environment temperature 2 to 8 °C |
| Distribution end date | 2019-01-31 |
Catalogue Numbers
| Catalogue number | Size | Status |
|---|---|---|
| 280550 | Withdrawn |
Same Product Code
| Product | Sold by | Catalogue numbers |
|---|---|---|
| GeneXpert Dx System | Cepheid | 37 |
| Alinity m | Abbott Molecular, Inc. | 20 |
| GeneXpert Infinity-80 System | Cepheid | 13 |
| Fluorescence Quantitative PCR Detection System | Anitoa Biotechnology(Hangzhou)Co., Ltd. | 12 |
| GeneXpert® | Cepheid | 10 |
| Certified Pre-Owned GeneXpert Dx System | Cepheid | 10 |
| Bioer GeneExplorer Thermal Cycler | Hangzhou Bioer Technology Co., Ltd. | 7 |
| m2000 | Abbott Molecular, Inc. | 5 |
| GeneXpert Infinity-48s System | Cepheid | 5 |
| Panther Fusion® | Hologic, Inc. | 4 |
| Bioer LineGene 9600 Plus Fluorescent Quantitative Detection System | Hangzhou Bioer Technology Co., Ltd. | 4 |
| Bioer QuantGene 9600 Fluorescent Quantitative Detection System | Hangzhou Bioer Technology Co., Ltd. | 3 |
Showing 12 of 188. All products under Real Time Nucleic Acid Amplification System
510(K) Clearances
| Clearance | Decision date |
|---|---|
| K243922 | 2025-03-20 |
| K222779 | 2023-01-26 |
| K220480 | 2022-07-11 |
Recalls
2 of these are Class II. Class II means a temporary or reversible health consequence. Volume alone is not a quality signal: a manufacturer that ships more recalls more.
| Initiated | FDA class | Root cause | Reason | Quantity |
|---|---|---|---|---|
| 2022-06-22 | Class II | Device Design | When a run is aborted, the cooling period protection does not occur and the user is able to open the lid prior to the completion of the cooling period, exposing the user to heated instrument components. Incidental contact with heated instru | 459 (394 US, 65 OUS) |
| 2022-02-15 | Class II | Component design/selection | The photomultiplier tube, may have compromised functionality in some instruments. As a result, the user is likely to observe an increase in false-positive or indeterminate results. | 452 units (398 US, 54 OUS) |
Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.