| Olympus Corporation of the Americas | 510(k) number | 2025-09-11 | Class I | Device Design | Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment. | 0 units US; 1,913 units OUS | |
| Olympus Corporation of the Americas | 510(k) number | 2025-09-11 | Class I | Device Design | Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment. | 4,297 units US; 4,290 units OUS | |
| Olympus Corporation of the Americas | 510(k) number | 2025-09-11 | Class I | Device Design | Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment. | 0 units US; 2,092 units OUS | |
| Olympus Corporation of the Americas | 510(k) number | 2025-09-11 | Class I | Device Design | Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment. | 0 units US; 1,867 units OUS | |
| Olympus Corporation of the Americas | 510(k) number | 2025-09-11 | Class I | Device Design | Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment. | 5,221 units US; 4,321 units OUS | |
| Olympus Corporation of the Americas | 510(k) number | 2025-09-11 | Class I | Device Design | Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment. | 3,046 units US; 2,414 units OUS | |
| Olympus Corporation of the Americas | 510(k) number | 2025-09-11 | Class I | Device Design | Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment. | 1 unit US; 2,306 units OUS | |
| Olympus Corporation of the Americas | 510(k) number | 2025-09-11 | Class I | Device Design | Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment. | 916 units US; 1,402 units OUS | |
| Olympus Corporation of the Americas | 510(k) number | 2025-09-11 | Class I | Device Design | Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment. | 361 units US; 258 units OUS | |
| Shirakawa Olympus Co., Ltd. | UDI-DI | 2023-12-04 | Class II | Under Investigation by firm | Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible. | 2600 US; 3214 OUS | |
| Shirakawa Olympus Co., Ltd. | UDI-DI | 2023-12-04 | Class II | Under Investigation by firm | Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible. | 1187 US; 1652 OUS | |
| Olympus Corporation of the Americas | 510(k) number | 2023-10-12 | Class I | Under Investigation by firm | There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope. | 32211 devices | |
| Aizu Olympus Co., Ltd. | 510(k) number | 2023-09-25 | Class II | Under Investigation by firm | Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead t | 1123 units | |
| Olympus Corporation of the Americas | 510(k) number | 2023-07-27 | Class II | Labeling Change Control | Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large. | 9449 units | |
| Olympus Corporation of the Americas | UDI-DI | 2023-07-27 | Class II | Labeling Change Control | Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large. | 235 units | |
| Olympus Corporation of the Americas | UDI-DI | 2023-07-27 | Class II | Labeling Change Control | Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large. | 3949 units | |
| Olympus Corporation of the Americas | 510(k) number | 2023-06-08 | Class I | Under Investigation by firm | There have been complaints of endobronchial combustion during therapeutic laser procedures with the bronchoscope. | 1 units | |
| Olympus Corporation of the Americas | 510(k) number | 2023-06-08 | Class I | Under Investigation by firm | There have been complaints of endobronchial combustion during therapeutic laser procedures with the bronchoscope. | 9,684 units | |
| Aizu Olympus Co., Ltd. | UDI-DI | 2022-07-20 | Class II | Under Investigation by firm | The recalled unit was utilized in a veterinary endoscopy procedure in advance of being assigned to healthcare facilities as a loaner. | | 2025-01-22 |