| 2024-09-30 | Class II | Process change control | Disposable biopsy instruments were potentially manufactured missing the backend notch retention feature on the needle that results in inadequate securing of the needle inside the device, which could lead to needle detachment and misplacemen | 71,498 | |
| 2023-11-08 | Class II | Process control | The internal diameter of the coaxial cannula may be smaller or larger than the external diameter of the biopsy needle, and the length of the cannula may exceed the stated length on the label. As a result, the biopsy needle may not fit prope | 1,243,937 | |
| 2023-08-10 | Class II | Device Design | Diameter mismatch between coaxial and biopsy needle (cutting cannula) in disposable core biopsy instrument kits, where the coaxial cannula inner diameter prohibits the biopsy needle (cutting cannula) from properly fitting into the coaxial c | 1,545 | |
| 2023-04-11 | Class II | Mixed-up of materials/components | Coaxial cannula internal diameter and length of biopsy instruments exceed the external diameter and length of the biopsy needle, prohibiting the biopsy needle from properly fitting into the coaxial cannula and accessing the target tissue, w | 4,120 | |
| 2023-02-28 | Class II | Mixed-up of materials/components | The external diameter of the biopsy instrument is larger than the internal diameter of the coaxial needle, which could lead to: 1) Procedural delay, 2) Requirement of an additional device, 3), Needle tip breakage, if used despite incompatib | 38,635 | |
| 2021-08-24 | Class II | Process control | Due to misbranded products shipped to customers. | 360 devices | 2024-12-23 |
| 2021-08-24 | Class II | Process control | Due to misbranded products shipped to customers. | 360 devices | 2024-12-23 |
| 2021-06-23 | Class II | Packaging process control | Due to complaints received regarding incomplete/open packaging seals potentially causing product to non-sterile. | 1,200 device | |
| 2021-06-23 | Class II | Packaging process control | Due to complaints received regarding incomplete/open packaging seals potentially causing product to non-sterile. | 560 devices | |
| 2021-06-23 | Class II | Packaging process control | Due to complaints received regarding incomplete/open packaging seals potentially causing product to non-sterile. | 3,961 devices | |
| 2021-01-14 | Class II | Process control | Due to the Penetration Depth Switch used to select the desired penetration depth of the biopsy needle for soft tissue biopsies, detaches from the device during use. | Total 131,876 kits (102,210 kits U.S. and 29,666 kits O.U.S. | 2025-01-28 |
| 2020-03-17 | Class II | Packaging process control | Sterility compromised due to packaging pouches that may not be sealed properly | 8,170 units (US: 420 and OUS: 7,750) | 2024-07-15 |
| 2020-01-03 | Class II | Component change control | Lot numbers and products inadvertently not included in the scope of the previous recall. Original recall was conducted due to an increase in probe failures for leaks, suction issues, and failure to obtain samples. | 1,585 units | 2022-03-17 |
| 2020-01-03 | Class II | Component change control | Lot numbers and products inadvertently not included in the scope of the previous recall. Original recall was conducted due to an increase in probe failures for leaks, suction issues, and failure to obtain samples. | 1,085 units | 2022-03-17 |
| 2020-01-03 | Class II | Component change control | Lot numbers and products inadvertently not included in the scope of the previous recall. Original recall was conducted due to an increase in probe failures for leaks, suction issues, and failure to obtain samples. | 2,050 units | 2022-03-17 |
| 2020-01-03 | Class II | Component change control | Lot numbers and products inadvertently not included in the scope of the previous recall. Original recall was conducted due to an increase in probe failures for leaks, suction issues, and failure to obtain samples. | 12,875 units | 2022-03-17 |
| 2019-05-02 | Class II | Component change control | There has been an increase in complaints related to leaks, suction issues, and failure to obtain samples while using the product. | 29,918 units | 2022-03-03 |
| 2019-05-02 | Class II | Component change control | There has been an increase in complaints related to leaks, suction issues, and failure to obtain samples while using the product. | 125,292 units | 2022-03-03 |
| 2019-05-02 | Class II | Component change control | There has been an increase in complaints related to leaks, suction issues, and failure to obtain samples while using the product. | 239 units | 2022-03-03 |
| 2019-05-02 | Class II | Component change control | There has been an increase in complaints related to leaks, suction issues, and failure to obtain samples while using the product. | 36,640 units | 2022-03-03 |
| 2019-05-02 | Class II | Component change control | There has been an increase in complaints related to leaks, suction issues, and failure to obtain samples while using the product. | 12,833 units | 2022-03-03 |
| 2019-05-02 | Class II | Component change control | There has been an increase in complaints related to leaks, suction issues, and failure to obtain samples while using the product. | 5,298 units | 2022-03-03 |
| 2017-09-22 | Class II | Device Design | During the review of May 2017 complaint data, a significant increase in MAX-CORE Disposable Core Biopsy Instrument complaints was observed for Failure to Prime, Failure to Fire, Failure to Obtain Sample, and Self-Activation. | 110,550 units | 2018-09-27 |
| 2015-04-09 | Class II | Process design | The firm is recalling Bard MAX-CORE Disposable Core Biopsy Instrument due to the risk of having self-activation-related issues. | 7,995 units | 2015-09-25 |
| 2015-03-17 | Class II | Nonconforming Material/Component | Bard Peripheral Vascular is recalling the Bard Monopty Disposable Core Biopsy Instruments and Kits because they may be at risk of having activation-related issues. | 29,342 units | 2015-08-03 |
| 2008-09-12 | Not classified | Device Design | Some of the Vacora Biopsy Vacuum Assisted Biopsy Probe thumb wheels from these lots may fracture when fired using the prime/pierce option. | 225 | 2012-09-19 |