| 2024-08-20 | Class I | Other | Baxter Healthcare Corporation has received increased customer reports of particulate matter in the Automated Compounding Device Inlets (disposable inlet), Product Codes H938173, H938174, H938175, H938176, used with the ExactaMix and ExactaM | 2,140,046 units | |
| 2024-08-20 | Class I | Other | Baxter Healthcare Corporation has received increased customer reports of particulate matter in the Automated Compounding Device Inlets (disposable inlet), Product Codes H938173, H938174, H938175, H938176, used with the ExactaMix and ExactaM | 3,580,283 units | |
| 2024-08-20 | Class I | Other | Baxter Healthcare Corporation has received increased customer reports of particulate matter in the Automated Compounding Device Inlets (disposable inlet), Product Codes H938173, H938174, H938175, H938176, used with the ExactaMix and ExactaM | 5,828,540 units | |
| 2024-08-20 | Class I | Other | Baxter Healthcare Corporation has received increased customer reports of particulate matter in the Automated Compounding Device Inlets (disposable inlet), Product Codes H938173, H938174, H938175, H938176, used with the ExactaMix and ExactaM | 1,548,782 units | |
| 2023-12-22 | Class I | Device Design | An error was identified in software versions 2.0.8 and 2.1.8 while using the “Use Some Overfill” feature which may lead to over-delivery of an ingredient. This issue can result in redundant ingredient delivery of the requested overfill volu | 67 devices | |
| 2023-12-22 | Class I | Device Design | An error was identified in software versions 2.0.8 and 2.1.8 while using the “Use Some Overfill” feature which may lead to over-delivery of an ingredient. This issue can result in redundant ingredient delivery of the requested overfill volu | 8 devices | |
| 2022-07-14 | Class II | Process control | There is a potential for leaking valves on ports 1 to 4 in certain lots of valve sets. | 448,960 units | |
| 2019-09-18 | Class II | Process control | Potential ability to leak once used for compounding. | 11,389,800 units total | 2021-02-16 |
| 2018-04-25 | Class II | Labeling Change Control | Specific lots of EXACTAMIX Inlets listed below due to incorrectly performed over-labeling. The inlets were over-labeled with a blank label, or with a label containing an extended expiration date. If the product is used beyond the expiratio | 98700 units | 2020-12-16 |
| 2018-04-25 | Class II | Labeling Change Control | Specific lots of EXACTAMIX Inlets listed below due to incorrectly performed over-labeling. The inlets were over-labeled with a blank label, or with a label containing an extended expiration date. If the product is used beyond the expiratio | 59475 units | 2020-12-16 |
| 2018-04-25 | Class II | Labeling Change Control | Specific lots of EXACTAMIX Inlets listed below due to incorrectly performed over-labeling. The inlets were over-labeled with a blank label, or with a label containing an extended expiration date. If the product is used beyond the expiratio | 75350 units | 2020-12-16 |
| 2018-04-25 | Class II | Labeling Change Control | Specific lots of EXACTAMIX Inlets listed below due to incorrectly performed over-labeling. The inlets were over-labeled with a blank label, or with a label containing an extended expiration date. If the product is used beyond the expiratio | 112975 units | 2020-12-16 |
| 2014-06-10 | Class II | Software design | If the universal ingredient (UI) in an active configuration is changed using the Configuration Editor, a flush of the outlet pump tube will not be initiated by the software. It could result in the original UI remaining in the tube which co | 1758 | 2017-02-22 |
| 2013-09-25 | Class II | Packaging process control | Incomplete foil seal on one lot of sterile product. | 38,400 units | 2014-07-22 |
| 2011-08-15 | Not classified | Software design | There is the potential for an order entry error during use of the LOGIX Order Entry (OE) software. When adding ingredients to a TPN order, the user can accept the pre-populated “per units” of “Bag” or select other “per units”. A dosing er | 95 units | 2017-01-05 |
| 2011-03-29 | Not classified | Software design | The LOGIX OE automated backup does not retain the three most recent backup files and instead will only retain the last three backups of the previous year. The software stores backup files in numeric order rather than chronological date ord | 84 units | 2017-01-05 |
| 2011-01-27 | Not classified | Other | The AUTOMIX compounder may not respond correctly to the key pressed by the operator using the keypad. The incorrect keypad response issue can occur both in manual data entry mode (STD Mode) and when entering order numbers to download from | 3100 units | 2012-02-16 |
| 2011-01-27 | Not classified | Other | The AUTOMIX compounder may not respond correctly to the key pressed by the operator using the keypad. The incorrect keypad response issue can occur both in manual data entry mode (STD Mode) and when entering order numbers to download from | 950 units | 2012-02-16 |
| 2011-01-27 | Not classified | Other | The AUTOMIX compounder may not respond correctly to the key pressed by the operator using the keypad. The incorrect keypad response issue can occur both in manual data entry mode (STD Mode) and when entering order numbers to download from | 840 units | 2012-02-16 |
| 2011-01-27 | Not classified | Other | The AUTOMIX compounder may not respond correctly to the key pressed by the operator using the keypad. The incorrect keypad response issue can occur both in manual data entry mode (STD Mode) and when entering order numbers to download from | 270 units | 2012-02-16 |
| 2008-03-28 | Not classified | Software design | Software anomaly results in inaccurate information being printed out on the Delivery Report, even though the compounder performed the compounding correctly. | 5 units | 2008-08-25 |
| 2004-12-20 | Not classified | Other | Logix-CM software version 3.0.4 has not completed its official release process. | 5 units | 2005-05-19 |
| 2004-12-08 | Not classified | Other | A software anomaly in the Logix CM software could result in a drug mixture being compounded without one of the ingredients if two drugs have the same indentifier in Logix and another order entry program being used with the Logix software. | 12 units | 2005-11-01 |