Regulatory record for RAPIDLab® 1200 Series CO-ox Sample Chamber
Listed by Siemens Healthcare Diagnostics Inc, United States
FDA product code GKR
Device type System, Hemoglobin, Automated
Device class Class 2
Regulation 21 CFR 864.5620
Establishment Contract manufacturer, Specification developer, Repacker / relabeller, Complaint-file establishment, Manufacturer, Exporter
510(k) K031560
Record published 2016-09-24
Record revised 2021-11-19 (version 5)

Specifications

Specifications
MRI safety Labeling does not contain MRI Safety Information
Label carries lot or batch number, expiry date

Catalogue Numbers

Catalogue numbers for RAPIDLab® 1200 Series CO-ox Sample Chamber
Catalogue number Size Status
10324165 In distribution

Same Product Code

Other products under the same FDA product code
Product Sold by Catalogue numbers
RAPIDPoint® 405/500 Measurement Cartridge Siemens Healthcare Diagnostics Inc 8
McKesson Mckesson Medical-Surgical Inc. 6
HemoCue® Hb 301 Microcuvettes Hemocue AB 5
HemoCue® Hb 201 Microcuvettes Hemocue AB 4
Henry Schein Henry Schein, Inc. 4
Mission Plus Hemoglobin (Hb) Monitoring System Acon Laboratories, Inc. 3
Mission Plus Hemoglobin (Hb) Test Cartridges Acon Laboratories, Inc. 3
DiaSpect Tm EKF Diagnostic Gmbh 2
HemoCue® Hb 201+ Analyzer Hemocue AB 2
Mission Capillary Transfer Tubes Acon Laboratories, Inc. 2
Mission Plus Hemoglobin (Hb) Optical Verifier Acon Laboratories, Inc. 2
HemoCue® Hb301 Analyzer Hemocue AB 2

Showing 12 of 43. All products under System, Hemoglobin, Automated

Recalls Naming This Product

14 recalls name this product by its own identifiers, the 510(k) it is cleared under.

Recalls naming the identifiers of RAPIDLab® 1200 Series CO-ox Sample Chamber
Recalled by Identified by Initiated FDA class Root cause Reason Quantity Closed
Siemens Healthcare Diagnostics Inc 510(k) number 2018-07-16 Class II Device Design Therapeutic levels (1mg/mL and 2mg/mL) of Hydroxocobalamin may interfere with tHb and some of the CO-Ox fractions that are reported on the RAPIDPoint and RAPIDLab Blood Gas systems indicated. This may cause lower than expected values for ca 138 2020-06-23
Siemens Healthcare Diagnostics Inc 510(k) number 2018-07-16 Class II Device Design Therapeutic levels (1mg/mL and 2mg/mL) of Hydroxocobalamin may interfere with tHb and some of the CO-Ox fractions that are reported on the RAPIDPoint and RAPIDLab Blood Gas systems indicated. This may cause lower than expected values for ca 2357 2020-06-23
Siemens Healthcare Diagnostics Inc 510(k) number 2017-04-10 Class II Software design Potential Patient Demographic Error with Blank Patient ID Field 159 units 2019-08-20
Siemens Healthcare Diagnostics Inc 510(k) number 2016-08-11 Class II Software design There is a potential for the first and/or last name of one patient to be printed with Patient ID and result data from a different patient, even though those fields have been turned Off for the system 144 units 2020-07-16
Siemens Healthcare Diagnostics Inc 510(k) number 2016-08-11 Class II Software design There is a potential for the first and/or last name of one patient to be printed with Patient ID and result data from a different patient, even though those fields have been turned Off for the system 197 units 2020-07-16
Siemens Healthcare Diagnostics Inc 510(k) number 2015-05-06 Class II Software design D50 and D51 Diagnostic error codes are not functional. 122 devices 2017-08-02
Siemens Healthcare Diagnostics Inc 510(k) number 2012-09-17 Class II Software Manufacturing/Software Deployment RAPIDPoint 400 Series, RAPIDPoint 500, or
RAPIDLab 1200 Series blood gas analyzer, users can inadvertently enter a wrong patient ID on the demographics screen, which ultimately could lead to a misidentification of the sample
179 units 2015-04-01
Siemens Healthcare Diagnostics Inc 510(k) number 2012-09-17 Class II Software Manufacturing/Software Deployment RAPIDPoint 400 Series, RAPIDPoint 500, or
RAPIDLab 1200 Series blood gas analyzer, users can inadvertently enter a wrong patient ID on the demographics screen, which ultimately could lead to a misidentification of the sample
274 units 2015-04-01
Siemens Healthcare Diagnostics Inc 510(k) number 2006-04-04 Not classified Software design Invalid CO-oximeter values; Bayer HealthCare determined that under very specific conditions, the RapidLab 1245 or RapidLab 1265 systems may report invalid CO-oximeter values to an LIS, Rapidlink, or Rapidcomm™ data management system. Domestic – 124 units, Foreign-364 units (all systems) 2012-07-30
Siemens Healthcare Diagnostics Inc 510(k) number 2006-04-04 Not classified Software design Invalid CO-oximeter values; Bayer HealthCare determined that under very specific conditions, the RapidLab 1245 or RapidLab 1265 systems may report invalid CO-oximeter values to an LIS, Rapidlink, or Rapidcomm™ data management system. Domestic – 124 units, Foreign-364 units (all systems) 2012-07-30
Siemens Healthcare Diagnostics Inc 510(k) number 2006-01-20 Not classified Other When stored data is recalled from memory, the results do not agree with the original results. Software anomaly: a calculation error occurs when using the correlation adjustment feature on the RapidLab 1200 Systems with software versions 1. 2012-05-02
Siemens Healthcare Diagnostics Inc 510(k) number 2006-01-20 Not classified Other When stored data is recalled from memory, the results do not agree with the original results. Software anomaly: a calculation error occurs when using the correlation adjustment feature on the RapidLab 1200 Systems with software versions 1. 2012-05-02
Siemens Healthcare Diagnostics Inc 510(k) number 2006-01-20 Not classified Other When stored data is recalled from memory, the results do not agree with the original results. Software anomaly: a calculation error occurs when using the correlation adjustment feature on the RapidLab 1200 Systems with software versions 1. 2012-05-02
Siemens Healthcare Diagnostics Inc 510(k) number 2006-01-20 Not classified Other When stored data is recalled from memory, the results do not agree with the original results. Software anomaly: a calculation error occurs when using the correlation adjustment feature on the RapidLab 1200 Systems with software versions 1. 2012-05-02

Recalls

2 of these are Class II. Class II means a temporary or reversible health consequence. Volume alone is not a quality signal: a manufacturer that ships more recalls more.

Recalls referencing this product code
Initiated FDA class Root cause Reason Quantity Closed
2018-07-16 Class II Device Design Therapeutic levels (1mg/mL and 2mg/mL) of Hydroxocobalamin may interfere with tHb and some of the CO-Ox fractions that are reported on the RAPIDPoint and RAPIDLab Blood Gas systems indicated. This may cause lower than expected values for ca 10199 2020-06-23
2018-07-16 Class II Device Design Therapeutic levels (1mg/mL and 2mg/mL) of Hydroxocobalamin may interfere with tHb and some of the CO-Ox fractions that are reported on the RAPIDPoint and RAPIDLab Blood Gas systems indicated. This may cause lower than expected values for ca 2248 2020-06-23

Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.

Where China Actually Supplies

Every device page names the company that lists it and the factories registered to make the same thing. These are the sections with the most of them.

5,100manufacturers registered in China
2,581device types they are registered for
340,248device records, each naming its maker
31provinces with a registered manufacturer