| 2026-04-24 | Class II | Under Investigation by firm | The cannulation of the CCS screw is not centered. | 13 units | |
| 2026-01-09 | Class II | Labeling mix-ups | Mix up between specified lots of 2.5mm and 2.8mm outer diameter screws. | 0 | |
| 2026-01-09 | Class II | Labeling mix-ups | Mix up between specified lots of 2.5mm and 2.8mm outer diameter screws. | 126 | |
| 2024-04-12 | Class II | Process control | Screw is 16mm long instead of 22mm long. | 110 units | |
| 2022-02-10 | Class II | Nonconforming Material/Component | Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect with and remove from the proximal Nextra implant, causing delay in surgery | 285 units | 2024-10-10 |
| 2022-02-10 | Class II | Nonconforming Material/Component | Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect with and remove from the proximal Nextra implant, causing delay in surgery | 239 units | 2024-10-10 |
| 2022-02-10 | Class II | Nonconforming Material/Component | Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect with and remove from the proximal Nextra implant, causing delay in surgery | 459 units | 2024-10-10 |
| 2022-02-10 | Class II | Nonconforming Material/Component | Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect with and remove from the proximal Nextra implant, causing delay in surgery | 4 units | 2024-10-10 |
| 2022-02-10 | Class II | Nonconforming Material/Component | Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect with and remove from the proximal Nextra implant, causing delay in surgery | 89 units | 2024-10-10 |
| 2022-01-13 | Class II | Device Design | During insertion of the middle phalanx implant into bone, the driver may deform the implant locking mechanism and result in inadequate mating force between the middle and proximal phalanx implants, resulting in surgical delay, in-situ remov | 23 units | 2024-05-24 |
| 2022-01-13 | Class II | Device Design | During insertion of the middle phalanx implant into bone, the driver may deform the implant locking mechanism and result in inadequate mating force between the middle and proximal phalanx implants, resulting in surgical delay, in-situ remov | 24 units | 2024-05-24 |
| 2021-03-26 | Class II | Under Investigation by firm | Product could potentially contain an undersized inner diameter of the 40mm drill bushing which result in not allowing the drill bit to pass through the drill bushing, causing inability of surgeon to complete surgical procedure as intended | 111 units | 2021-11-10 |
| 2021-03-26 | Class II | Under Investigation by firm | Product could potentially contain an undersized inner diameter of the 40mm drill bushing which result in not allowing the drill bit to pass through the drill bushing, causing inability of surgeon to complete surgical procedure as intended | 123 units | 2021-11-10 |
| 2020-02-12 | Class II | Under Investigation by firm | Proximal end of the driver in the affected product may be oversized in varying degrees, potentially making it difficult to connect with and remove from the proximal Nextra implant | 190 units | 2020-12-17 |
| 2019-11-11 | Class II | Packaging process control | Components 2mm shorter than indicated on the package label | 86 units | 2020-09-10 |
| 2019-11-11 | Class II | Packaging process control | Components 2mm shorter than indicated on the package label | 86 units | 2020-09-10 |
| 2019-11-11 | Class II | Packaging process control | Components 2mm shorter than indicated on the package label | 87 units | 2020-09-10 |
| 2014-10-01 | Class II | Process design | A recall of the APTUS Ulna Shortening 2.5 instrument was initiated due to three (3) complaints of breakage of the temporary tension bolt. | 44 | 2015-08-06 |