Toshiba Scanner

This device was previously sold by TOSHIBA MEDICAL SYSTEMS CORPORATION under the brand name TOSHIBA SCANNER. Effective Jan 2018, the company name changed to CANON MEDICAL SYSTEMS…

Registry build 21.09.26

This device was previously sold by TOSHIBA MEDICAL SYSTEMS CORPORATION under the brand name TOSHIBA SCANNER. Effective Jan 2018, the company name changed to CANON MEDICAL SYSTEMS CORPORATION. The device brand name is now CT SCANNER.

8manufacturers in China registered for this device type
10catalogue numbers in this listing

Who Else Makes It

Concentrated in Beijing (2), Shanghai (2), Jiangsu (2). 2 of them are registered as contract manufacturers, able to build to a buyer’s own specification.

Chinese manufacturers registered for FDA product code JAK
Manufacturer City Province Registered as Device types
Philips Healthcare (Suzhou) Co., Ltd. Suzhou Jiangsu Contract manufacturer 5
Qisda Electronics (Suzhou) Co., Ltd. Suzhou Jiangsu Contract manufacturer 5
GE Hangwei Medical Systems Co., Ltd. Beijing Beijing Specification developer 22
GE Healthcare (China) Co., Ltd. Beijing Beijing Specification developer 22
Shanghai United Imaging Healthcare Co., Ltd. Shanghai Shanghai Manufacturer 12
Siemens Shanghai, Medical Equipment Ltd. Shanghai Shanghai Contract manufacturer 10
Neusoft Medical Systems Co., Ltd. Shenyang Liaoning Manufacturer 1
Philips Health Technology (China) Co., Ltd. Shenyang Branch Shenyang Liaoning Specification developer 1

Record

Regulatory record for Toshiba Scanner
Listed by Canon Inc., Japan
FDA product code JAK
Device type System, X-Ray, Tomography, Computed
Device class Class 2
Regulation 21 CFR 892.1750
Establishment Contract manufacturer, Specification developer, Manufacturer
510(k) K151833
Record published 2018-03-02
Record revised 2026-03-20 (version 5)

Specifications

Specifications
MRI safety Labeling does not contain MRI Safety Information
Label carries serial number

Catalogue Numbers

Catalogue numbers for Toshiba Scanner
Catalogue number Size Status
TSX-035A In distribution
TSX-201A In distribution
TSX-036A In distribution
TSX-301C In distribution
TSX-302A In distribution
TSX-305A In distribution
TSX-101A In distribution
TSX-303A In distribution
TSX-303B In distribution
TSX-301A In distribution

Same Product Code

Other products under the same FDA product code
Product Sold by Catalogue numbers
Visicoil IZI Medical Products LLC 48
Visicoil MR IZI Medical Products LLC 41
Illumisite Covidien LP 25
Imbio LDA Imbio, LLC 24
Edge Covidien LP 20
Revolution GE Hangwei Medical Systems Co., Ltd. 20
Gold Fiducial IZI Medical Products LLC 20
superDimension Covidien LP 17
Lodox Lodox Systems (PTY) Ltd. 12
Revolution GE Medical Systems, L.L.C. 12
syngo.CT Dual Energy Siemens Healthcare Limited 7
Imactis® Imactis, SAS 6

Showing 12 of 495. All products under System, X-Ray, Tomography, Computed

Recalls Naming This Product

29 recalls name this product by its own identifiers, the 510(k) it is cleared under.

Recalls naming the identifiers of Toshiba Scanner
Recalled by Identified by Initiated FDA class Root cause Reason Quantity Closed
Canon Medical System, USA, Inc. 510(k) number 2020-11-09 Class II Software design A software problem has been identified which could result in the CT Scanner not proceeding to the next actual scan even though automatic start of the next scan is specified. This could result in the CT scan failing resulting in rescanning 60 units 2022-12-16
Canon Medical System, USA, Inc. 510(k) number 2019-03-25 Class II Software design Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would oc 6 devices
Canon Medical System, USA, Inc. 510(k) number 2019-03-25 Class II Software design Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would oc 2 devices
Canon Medical System, USA, Inc. 510(k) number 2019-03-25 Class II Software design Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would oc 1 device
Canon Medical System, USA, Inc. 510(k) number 2019-03-25 Class II Software design Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would oc 10 devices
Canon Medical System, USA, Inc. 510(k) number 2019-03-25 Class II Software design Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would oc 10 devices
Canon Medical System, USA, Inc. 510(k) number 2019-03-25 Class II Software design Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would oc 35 devices
Toshiba American Medical Systems Inc 510(k) number 2016-10-25 Class II Device Design Toshiba American Medical Systems (TAMS) is recalling the Aquilion CT system because it
has been reported that the cooling oil in the x-ray tube heat exchanger may leak during system operation.
291 units 2017-08-19
Toshiba American Medical Systems Inc 510(k) number 2016-04-12 Class II Device Design Toshiba American Medical Systems, Inc. is recalling the TSX-035A because of a potential problem with the Aquilion Lightning CT System iStation cover falling off of the gantry. 1 unit 2016-05-19
Toshiba American Medical Systems Inc 510(k) number 2015-05-08 Class II Radiation Control for Health and Safety Act If two specific operations are performed in multi-phase helical scanning, the acquired raw data may not be saved. 459 2016-10-26
Toshiba American Medical Systems Inc 510(k) number 2015-05-08 Class II Radiation Control for Health and Safety Act If two specific operations are performed in multi-phase helical scanning, the acquired raw data may not be saved. 459 2016-10-26
Toshiba American Medical Systems Inc 510(k) number 2015-05-08 Class II Radiation Control for Health and Safety Act If two specific operations are performed in multi-phase helical scanning, the acquired raw data may not be saved. 459 2016-10-26
Toshiba American Medical Systems Inc 510(k) number 2015-05-08 Class II Radiation Control for Health and Safety Act If two specific operations are performed in multi-phase helical scanning, the acquired raw data may not be saved. 459 2016-10-26
Toshiba American Medical Systems Inc 510(k) number 2015-05-08 Class II Radiation Control for Health and Safety Act If two specific operations are performed in multi-phase helical scanning, the acquired raw data may not be saved. 459 2016-10-26
Toshiba American Medical Systems Inc 510(k) number 2014-11-05 Class II Software design Potential problem with the cardiac function analysis software (CFA). Incorrect analysis results may be displayed in the Function Parameters for the Entire Heart displayed as analysis results of the CFA and in a Left-Ventricular Volume Curv 1 unit within the U.S. only 2015-10-09
Toshiba American Medical Systems Inc 510(k) number 2014-11-05 Class II Software design Potential problem with the cardiac function analysis software (CFA). Incorrect analysis results may be displayed in the Function Parameters for the Entire Heart displayed as analysis results of the CFA and in a Left-Ventricular Volume Curv 2 units within the U.S. only 2015-10-09
Toshiba American Medical Systems Inc 510(k) number 2013-12-11 Class II Other Toshiba America Medical Systems, Inc. initiated this recall due to the failure of the parts used in the X-ray high-voltage generator. In some cases, X-ray exposure cannot be performed. 8 units 2014-10-16
Toshiba American Medical Systems Inc 510(k) number 2013-10-01 Class II Software design Toshiba intitiated this recall because their investigation revealed that due to a problem of the X-ray output control software, the X-ray output conditions may not be set correctly. As a result, X-ray exposure under the scan conditions used 338 units 2015-03-03
Toshiba American Medical Systems Inc 510(k) number 2012-11-14 Class II Process design Due to a problem with the control software of the x-ray high-voltage generator, if scanning is performed with a tube current less than 30mA, the control software may detect an error and perform error processing even though the x-ray output 1 unit 2013-04-08
Toshiba American Medical Systems Inc 510(k) number 2011-06-22 Not classified Software design Toshiba America Medical Systems is initiating a field correction on their TSX-301B: AQUILION Premium; System, X-Ray, Tomography, Computed,
TSX-301A: AQUILION One; System, X-Ray, Tomography, Computed,
TSX-20IA: AQUlLlON LB; System, X-Ray, To
155 units total 2014-08-11
Toshiba American Medical Systems Inc 510(k) number 2010-12-08 Not classified Software design There is the potential that the current may not be controlled correctly in examinations using the ECG modulation function, which adjusts the x-ray output in synchronization with the ECG waveform. 528 units 2013-03-18
Toshiba American Medical Systems Inc 510(k) number 2010-04-09 Not classified Software design The recall was initiated by Toshiba America Medical Systems (TAMS) because the software associated with TSX-101A: Aquilion 64 can prevent scanning from being performed at the specified helical pitch (couch movement speed). 35 2012-08-17
Toshiba American Medical Systems Inc 510(k) number 2009-07-13 Class II Software design 1. It has been found that in raw data processing for Go&Return helical scanning, when the raw data acquired in the first scan (Go) and that acquired in the second scan (Return) are selected and reconstructed at one time, incorrect couch pos 24 Systems 2012-09-04
Toshiba American Medical Systems Inc 510(k) number 2009-02-03 Not classified Software design ECG Gated Dynamic Volume Prospective scan may stop if preset eXam Plan is confirmed and CTA On is changed to CTA OFF, due to exposure control PWB. 16 2011-09-02
Toshiba American Medical Systems Inc 510(k) number 2009-01-08 Not classified Component design/selection Toshiba America Medical Systems Inc (TAMS) has initiated a field correction on the Aquilion 32 CT Scanners due to the potential for the bolus tracking software (SureStart) to cause delay of the helical scan, up to 10 seconds, when large vol 94 2012-06-01
Toshiba American Medical Systems Inc 510(k) number 2009-01-08 Not classified Component design/selection Toshiba America Medical Systems Inc (TAMS) has initiated a field correction on the Aquilion 32 CT Scanners due to the potential for the bolus tracking software (SureStart) to cause delay of the helical scan, up to 10 seconds, when large vol 397 2012-06-01
Toshiba American Medical Systems Inc 510(k) number 2009-01-08 Not classified Component design/selection Toshiba America Medical Systems Inc (TAMS) has initiated a field correction on the Aquilion 32 CT Scanners due to the potential for the bolus tracking software (SureStart) to cause delay of the helical scan, up to 10 seconds, when large vol 9 2012-06-01
Toshiba American Medical Systems Inc 510(k) number 2009-01-08 Not classified Component design/selection Toshiba America Medical Systems Inc (TAMS) has initiated a field correction on the Aquilion 32 CT Scanners due to the potential for the bolus tracking software (SureStart) to cause delay of the helical scan, up to 10 seconds, when large vol 47 2012-06-01
Toshiba American Medical Systems Inc 510(k) number 2008-05-13 Class II Software design The SureStart function may not operate in the usual manner, and you may obtain
abnormal analysis results when using the cerebral blood flow analysis function.
3 units 2012-09-27

Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.

Where China Actually Supplies

Every device page names the company that lists it and the factories registered to make the same thing. These are the sections with the most of them.

5,100manufacturers registered in China
2,581device types they are registered for
340,248device records, each naming its maker
31provinces with a registered manufacturer