Neusoft Medical Systems Co., Ltd.

Manufacturer and brand owner in Shenyang, Liaoning, China. 1 device type, 5 products catalogued, 23 clearances.

Registry build 21.09.26

Neusoft Medical Systems Co., Ltd. is a manufacturer holding a current FDA establishment registration. The registration covers System, X-Ray, Tomography, Computed. 5 of its products are catalogued under its own name. Its earliest product listing with the FDA dates from 2012.

1device type
5products catalogued under its own name
23510(k) clearances
17recalls

Registration

FDA registration record for Neusoft Medical Systems Co., Ltd.
Registered name NEUSOFT MEDICAL SYSTEMS CO., LTD.
Status Registered
Registration expires 2026
Registration number 3003575350
FEI number 3003575350
DUNS 724421537
Establishment type Manufacturer

Chinese Registration

Chinese company register details for Neusoft Medical Systems Co., Ltd.
Chinese legal name 东软医疗系统股份有限公司
Unified social credit code 91210112702087542R
Source NMPA UDI database

Location

Address of Neusoft Medical Systems Co., Ltd.
City Shenyang
Province Liaoning
Address No.177-1 Chuangxin Road
Postal code 110167
Country China

What It Makes

Device types Neusoft Medical Systems Co., Ltd. is registered to make
Device type Class Listed under the names Listed since Chinese makers
System, X-Ray, Tomography, Computed Class 2 NeuViz 64 In, NeuViz 64 En, NeuViz 64i,NeuViz 64e, ClearView, NeuViz 16 Essence Multi-Slice CT Scanner System, NeuViz 128 2012 8

Own-Brand Listings

Clearances

510(k) clearances naming Neusoft Medical Systems Co., Ltd. or its listings
Clearance Decision date Product code Route Held by
K230220 2023-10-11 JAK Traditional This company
K171198 2017-12-07 JAK Traditional This company
K171201 2017-09-13 JAK Traditional This company
K151383 2015-11-04 JAK Traditional This company
K151403 2015-11-02 KPS Traditional This company
K133373 2014-12-03 JAK Traditional This company
K121792 2012-11-16 JAK Traditional This company
K110077 2011-01-28 LLZ Traditional This company
K102435 2010-10-15 LLZ Traditional This company
K092742 2009-09-22 JAK Traditional This company
K092237 2009-09-09 LNH Traditional This company
K092235 2009-08-06 LLZ Traditional This company
K090552 2009-03-16 LLZ Traditional This company
K090173 2009-02-05 JAK Traditional This company
K083877 2009-01-13 ITX Traditional This company
K082485 2008-09-09 LNH Traditional This company
K080540 2008-03-13 JAK Traditional This company
K072841 2007-10-16 JAK Traditional This company
K071308 2007-08-16 JAK Traditional This company
K071154 2007-05-10 LNH Traditional This company
K062860 2006-10-27 LNH Special This company
K062451 2006-09-28 JAK Traditional This company
K061132 2006-05-24 LNH Special This company

Recalls

13 of these are Class II. Class II means a temporary or reversible health consequence. Volume alone is not a quality signal: a manufacturer that ships more recalls more.

Recalls naming Neusoft Medical Systems Co., Ltd.
Initiated FDA class Root cause Reason Quantity Closed
2018-10-22 Class II Software design Software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan,
there will be some electromagnetic interference in the system which may affect the control signal of the flying focal spot, r
6 2022-08-25
2018-10-22 Class II Software design Software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan,
there will be some electromagnetic interference in the system which may affect the control signal of the flying focal spot, r
37 2022-08-25
2018-10-22 Class II Software design Software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan,
there will be some electromagnetic interference in the system which may affect the control signal of the flying focal spot, r
2 2022-08-25
2017-05-19 Class II Software design Software defect 234 2020-11-19
2017-05-19 Class II Software design Software defect 154 2020-11-19
2016-08-31 Class II Software design The Firm discovered during contrast agent tracking scan, when the expected concentration of contrast agent is reached, later planned scan may be interrupted. The system must be powered down and restarted to clear the interruption error befo 23 units 2018-01-04
2016-08-31 Class II Software design The Firm discovered during contrast agent tracking scan, when the expected concentration of contrast agent is reached, later planned scan may be interrupted. The system must be powered down and restarted to clear the interruption error befo 23 units 2018-01-04
2016-03-24 Class II Under Investigation by firm The following issues are found in NeuViz 16 systems with software version 1.1.4.21425 and version 1.1.4.21426: 1) During the filming operation on MX 16-slice console software, the clipboard used for copying and pasting images is not cleared 37 units 2016-11-23
2015-05-02 Class II Device Design It was found by R&D in April, 2014 that there was a defect in NeuViz 64 system software version 1.0.5+P09. The icons of side decubitus didn’t meet the patient position description, when the scanner position was set to “Right” in the system 7 units 2016-11-04
2013-03-04 Class II Use error User error of the NeuViz 16 or NeuViz Dual may cause personnel injuries. There was a report where a finger was pinched when the table was moved backward while the tabletop was controlled by another person from the gantry left control pane 25 units 2013-11-22
2013-03-04 Class II Use error User error of the NeuViz 16 or NeuViz Dual may cause personnel injuries. There was a report where a finger was pinched when the table was moved backward while the tabletop was controlled by another person from the gantry left control pane 23 units 2013-11-22
2012-10-18 Class II Device Design The four screws that secure the patient table top to the carrier pulled out and caused the table top to detach. This issue may affect NeuViz Dual and NeuViz 16 scanners. 20 units of NeuViz 16 CT Scanner Systems and 25 units of Neu 2013-07-08
2012-06-08 Class II Device Design One of the four screws that secure the Oil-accumulator to the X-Ray Tube pulled out in the NeuViz 16 CT scanner, which caused the Oil-accumulator to detach. 15 units distributed in the US 2012-09-17
2011-02-09 Not classified Other Dimension measurement displays on a combined image were found inaccurate after the clinician merged multiple image series into one image. 9 units distributed in US 2011-08-24
2010-02-10 Not classified Other Potential for the CT patient table top to become detached from the carrier due to missing washers. 26 devices 2011-04-15
2008-06-23 Not classified Nonconforming Material/Component Potential for R-host box and components inside to be detached of its mounting during the gantry rotor rotation. There is a potential for components to be damaged inside the system gantry and for the internal components to be expelled from 15 units in US 2010-09-24
2007-12-17 Not classified Software design Labeling of patient position on scanned image does not match actual patient orientation. 15 units 2012-02-16

From the Company

Published by the company on its own website, not by the FDA.

Information published by Neusoft Medical Systems Co., Ltd.
Main products CT systems, MRI systems, DSA systems, DR systems, PET/CT systems
Certifications FDA, CE
Export markets 130+ countries and regions

Nearby Manufacturers

Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.

The manufacturer named here can complete this record.