NeuViz 128 Multi-Slice CT Scanner System

NeuViz 128 is a whole body X-ray CT scanner that produces cross sectional images from transmission measurements acquired around the patient. It combines a gantry, patient couch, operator…

Registry build 21.09.26

NeuViz 128 is a whole body X-ray CT scanner that produces cross sectional images from transmission measurements acquired around the patient. It combines a gantry, patient couch, operator console and image acquisition and reconstruction hardware and software.

An X-ray tube and detector array rotate continuously around the patient using fan beam geometry. The computer reconstructs images from measurements taken at different angles during axial or spiral scanning. A 64 row detector supports generation of up to 128 slices per rotation.

Image Reconstruction

The scanner provides filtered back projection and ClearView iterative reconstruction. Optional ClearInfinity reconstruction uses a trained neural network for image noise reduction. The gantry has a 72 cm opening and a standard reconstruction field of view up to 50 cm.

8manufacturers in China registered for this device type
2catalogue numbers in this listing
12510(k) clearances naming this product code
17recalls on this code

Who Else Makes It

Concentrated in Beijing (2), Shanghai (2), Jiangsu (2). 2 of them are registered as contract manufacturers, able to build to a buyer’s own specification.

Chinese manufacturers registered for FDA product code JAK
Manufacturer City Province Registered as Device types
Philips Healthcare (Suzhou) Co., Ltd. Suzhou Jiangsu Contract manufacturer 5
Qisda Electronics (Suzhou) Co., Ltd. Suzhou Jiangsu Contract manufacturer 5
GE Hangwei Medical Systems Co., Ltd. Beijing Beijing Specification developer 22
GE Healthcare (China) Co., Ltd. Beijing Beijing Specification developer 22
Shanghai United Imaging Healthcare Co., Ltd. Shanghai Shanghai Manufacturer 12
Siemens Shanghai, Medical Equipment Ltd. Shanghai Shanghai Contract manufacturer 10
Neusoft Medical Systems Co., Ltd. Shenyang Liaoning Manufacturer 1
Philips Health Technology (China) Co., Ltd. Shenyang Branch Shenyang Liaoning Specification developer 1

Record

Regulatory record for NeuViz 128 Multi-Slice CT Scanner System
Listed by Neusoft Medical Systems Co., Ltd., China
FDA product code JAK
Device type System, X-Ray, Tomography, Computed
Device class Class 2
Regulation 21 CFR 892.1750
Establishment Manufacturer
510(k) K230220
Record published 2023-10-23
Record revised 2023-06-29 (version 6)

Specifications

Specifications
MRI safety Labeling does not contain MRI Safety Information
Label carries manufacturing date, serial number

Catalogue Numbers

Catalogue numbers for NeuViz 128 Multi-Slice CT Scanner System
Catalogue number Size Status
NeuViz 128 In distribution
NeuViz 128 In distribution

Same Product Code

Other products under the same FDA product code
Product Sold by Catalogue numbers
Visicoil IZI Medical Products LLC 48
Visicoil MR IZI Medical Products LLC 41
Illumisite Covidien LP 25
Imbio LDA Imbio, LLC 24
Edge Covidien LP 20
Revolution GE Hangwei Medical Systems Co., Ltd. 20
Gold Fiducial IZI Medical Products LLC 20
superDimension Covidien LP 17
Lodox Lodox Systems (PTY) Ltd. 12
Revolution GE Medical Systems, L.L.C. 12
Toshiba Scanner Canon Inc. 10
syngo.CT Dual Energy Siemens Healthcare Limited 7

Showing 12 of 495. All products under System, X-Ray, Tomography, Computed

510(K) Clearances

FDA 510(k) clearances naming this product code
Clearance Decision date
K230220 2023-10-11
K171198 2017-12-07
K171201 2017-09-13
K151383 2015-11-04
K133373 2014-12-03
K121792 2012-11-16
K092742 2009-09-22
K090173 2009-02-05
K080540 2008-03-13
K072841 2007-10-16
K071308 2007-08-16
K062451 2006-09-28

Recalls Naming This Product

3 recalls name this product by its own identifiers, the 510(k) it is cleared under.

Recalls naming the identifiers of NeuViz 128 Multi-Slice CT Scanner System
Recalled by Identified by Initiated FDA class Root cause Reason Quantity Closed
Neusoft Medical Systems IMP & EX 510(k) number 2019-01-09 Class II Software design For surview scan length more than 500mm,if the user aborts or skips the reconstructed surview, the image may have the wrong field of view (FOV). When user plans and performs subsequent clinical scan with the wrong FOV the actual start and 3 2022-08-25
Neusoft Medical Systems Co., Ltd. 510(k) number 2018-10-22 Class II Software design Software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan,
there will be some electromagnetic interference in the system which may affect the control signal of the flying focal spot, r
6 2022-08-25
Neusoft Medical Systems Co., Ltd. 510(k) number 2017-05-19 Class II Software design Software defect 154 2020-11-19

Recalls

11 of these are Class II. Class II means a temporary or reversible health consequence. Volume alone is not a quality signal: a manufacturer that ships more recalls more.

Recalls referencing this product code
Initiated FDA class Root cause Reason Quantity Closed
2018-10-22 Class II Software design Software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan,
there will be some electromagnetic interference in the system which may affect the control signal of the flying focal spot, r
37 2022-08-25
2018-10-22 Class II Software design Software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan,
there will be some electromagnetic interference in the system which may affect the control signal of the flying focal spot, r
2 2022-08-25
2017-05-19 Class II Software design Software defect 234 2020-11-19
2016-08-31 Class II Software design The Firm discovered during contrast agent tracking scan, when the expected concentration of contrast agent is reached, later planned scan may be interrupted. The system must be powered down and restarted to clear the interruption error befo 23 units 2018-01-04
2016-08-31 Class II Software design The Firm discovered during contrast agent tracking scan, when the expected concentration of contrast agent is reached, later planned scan may be interrupted. The system must be powered down and restarted to clear the interruption error befo 23 units 2018-01-04
2016-03-24 Class II Under Investigation by firm The following issues are found in NeuViz 16 systems with software version 1.1.4.21425 and version 1.1.4.21426: 1) During the filming operation on MX 16-slice console software, the clipboard used for copying and pasting images is not cleared 37 units 2016-11-23
2015-05-02 Class II Device Design It was found by R&D in April, 2014 that there was a defect in NeuViz 64 system software version 1.0.5+P09. The icons of side decubitus didn’t meet the patient position description, when the scanner position was set to “Right” in the system 7 units 2016-11-04
2013-03-04 Class II Use error User error of the NeuViz 16 or NeuViz Dual may cause personnel injuries. There was a report where a finger was pinched when the table was moved backward while the tabletop was controlled by another person from the gantry left control pane 25 units 2013-11-22
2013-03-04 Class II Use error User error of the NeuViz 16 or NeuViz Dual may cause personnel injuries. There was a report where a finger was pinched when the table was moved backward while the tabletop was controlled by another person from the gantry left control pane 23 units 2013-11-22
2012-10-18 Class II Device Design The four screws that secure the patient table top to the carrier pulled out and caused the table top to detach. This issue may affect NeuViz Dual and NeuViz 16 scanners. 20 units of NeuViz 16 CT Scanner Systems and 25 units of Neu 2013-07-08
2012-06-08 Class II Device Design One of the four screws that secure the Oil-accumulator to the X-Ray Tube pulled out in the NeuViz 16 CT scanner, which caused the Oil-accumulator to detach. 15 units distributed in the US 2012-09-17
2011-02-09 Not classified Other Dimension measurement displays on a combined image were found inaccurate after the clinician merged multiple image series into one image. 9 units distributed in US 2011-08-24
2010-02-10 Not classified Other Potential for the CT patient table top to become detached from the carrier due to missing washers. 26 devices 2011-04-15
2008-06-23 Not classified Nonconforming Material/Component Potential for R-host box and components inside to be detached of its mounting during the gantry rotor rotation. There is a potential for components to be damaged inside the system gantry and for the internal components to be expelled from 15 units in US 2010-09-24
2007-12-17 Not classified Software design Labeling of patient position on scanned image does not match actual patient orientation. 15 units 2012-02-16

Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.

Where China Actually Supplies

Every device page names the company that lists it and the factories registered to make the same thing. These are the sections with the most of them.

5,100manufacturers registered in China
2,581device types they are registered for
340,248device records, each naming its maker
31provinces with a registered manufacturer