| Siemens Medical Solutions USA, Inc. | 510(k) number | 2017-02-20 | Class II | Labeling Change Control | The mobile-specific instructions was not shipped to a limited population of Biograph Mobile systems. | 18 units | 2017-09-17 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2017-02-20 | Class II | Labeling Change Control | The mobile-specific instructions was not shipped to a limited population of Biograph Mobile systems. | 6 units | 2017-09-17 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2017-02-20 | Class II | Labeling Change Control | The mobile-specific instructions was not shipped to a limited population of Biograph Mobile systems. | 9 units | 2017-09-17 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2017-02-20 | Class II | Labeling Change Control | The mobile-specific instructions was not shipped to a limited population of Biograph Mobile systems. | 4 units | 2017-09-17 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | 83 units | 2017-05-08 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | 36 units | 2017-05-08 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | 5 units | 2017-05-08 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | 48 units | 2017-05-08 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | 46 units | 2017-05-08 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | 1 unit | 2017-05-08 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | 1 unit | 2017-05-08 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | 102 units | 2017-05-08 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | 15 units | 2017-05-08 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | 16 units | 2017-05-08 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | 49 units | 2017-05-08 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | | 2017-05-08 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | 2 units | 2017-05-08 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | 4 units | 2017-05-08 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | 2 units | 2017-05-08 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | 9 units | 2017-05-08 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | 28 units | 2017-05-08 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | 3 units | 2017-05-08 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | 9 units | 2017-05-08 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | 12 units | 2017-05-08 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | 25 units | 2017-05-08 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | 2 units | 2017-05-08 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | 13 units | 2017-05-08 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | 143 units | 2017-05-08 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | 5 units | 2017-05-08 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | 47 units | 2017-05-08 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | 100 units | 2017-05-08 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | 51 units | 2017-05-08 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | 127 units | 2017-05-08 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | 22 units | 2017-05-08 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | 4 units | 2017-05-08 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | 2 units | 2017-05-08 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-04-22 | Class II | Nonconforming Material/Component | Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. | 63 units | 2017-05-08 |