Optima

Optima IGS 320 001 UDI

Registry build 21.09.26
5manufacturers in China registered for this device type
1catalogue number
1510(k) clearance

Who Else Makes It

Concentrated in Beijing (2), Shanghai (2), Guangdong (1).

Chinese manufacturers registered for FDA product code OWB
Manufacturer City Province Registered as Device types
GE Hangwei Medical Systems Co., Ltd. Beijing Beijing Specification developer 22
GE Healthcare (China) Co., Ltd. Beijing Beijing Specification developer 22
Shanghai United Imaging Healthcare Co., Ltd. Shanghai Shanghai Manufacturer 12
Siemens Shanghai, Medical Equipment Ltd. Shanghai Shanghai Contract manufacturer 10
Siemens Shenzhen Magnetic Resonance Ltd. Shenzhen Guangdong Contract manufacturer 8

Record

Regulatory record for Optima
Listed by Ge Hualun Medical Systems Co., Ltd., China
FDA product code OWB
Device type Interventional Fluoroscopic X-Ray System
Device class Class 2
Regulation 21 CFR 892.1650
510(k) K122457
Record published 2016-08-01
Record revised 2025-04-07 (version 6)

Specifications

Specifications
MRI safety Labeling does not contain MRI Safety Information
Prescription Prescription only
Label carries manufacturing date, serial number

Catalogue Numbers

Catalogue numbers for Optima
Catalogue number Size Status
IGS 320 001 In distribution

Same Product Code

Other products under the same FDA product code
Product Sold by Catalogue numbers
TrackX Trackx Technology Inc. 86
Trinias Shimadzu Corp. 54
Allura Xper FD Philips Medical Systems Nederland B.V. 21
Azurion Philips Medical Systems Nederland B.V. 19
LessRay Nuvasive, Inc. 16
Intelli-C Accessory Nrt – Nordisk Røntgen Teknik A/S 13
Pulse Nuvasive, Inc. 12
O-ARM® Medtronic Navigation, Inc. 11
Genoray Genoray Co., Ltd. 10
Fujifilm Healthcare Americas Corporation A.T.S. Applicazione Tecnologie Speciali SRL 7
Fujifilm Medical Systems U.S.A. Inc. A.T.S. Applicazione Tecnologie Speciali SRL 6
Stephanix A.T.S. Applicazione Tecnologie Speciali SRL 6

Showing 12 of 122. All products under Interventional Fluoroscopic X-Ray System

510(K) Clearances

FDA 510(k) clearances naming this product code
Clearance Decision date
K172700 2017-11-09

Recalls Naming This Product

20 recalls name this product by its own identifiers, the 510(k) it is cleared under.

Recalls naming the identifiers of Optima
Recalled by Identified by Initiated FDA class Root cause Reason Quantity Closed
GE Medical Systems, LLC 510(k) number 2019-03-15 Class II Device Design There is a potential for loss of the x-ray imaging function when the user changes field of view (FOV) from 30cm to 20cm or from 20cm to 30cm while releasing the fluoroscopy footswitch pedal simultaneously. 49 devices
GE Medical Systems SCS 510(k) number 2018-03-29 Class II Process control The Discovery IGS and Innova IGS systems may experience X-ray abort errors which could result in the loss of imaging mode before or during fluoroscopic use. 102 devices total 2021-06-15
GE Medical Systems SCS 510(k) number 2018-03-29 Class II Process control The Discovery IGS and Innova IGS systems may experience X-ray abort errors which could result in the loss of imaging mode before or during fluoroscopic use. 102 devices total 2021-06-15
GE Medical Systems, LLC 510(k) number 2015-12-18 Class II Nonconforming Material/Component GE Healthcare has recently become aware of a reported incident in which a Mavig suspension arm with monitor assembly fell to the floor. 2,111 (1078 US; 1033 OUS) 2016-09-07
GE Medical Systems, LLC 510(k) number 2015-11-13 Class II Nonconforming Material/Component GE Healthcare has recently become aware of a potential safety issue related to a non-recoverable loss of displayed imaging (loss of monitor video) involving Optima CL323i & Optima IGS 320 systems. 108 2016-09-07
GE Medical Systems, LLC 510(k) number 2015-11-13 Class II Nonconforming Material/Component GE Healthcare has recently become aware of a potential safety issue related to a non-recoverable loss of displayed imaging (loss of monitor video) involving Optima CL323i & Optima IGS 320 systems. 45 2016-09-07
GE Medical Systems, LLC 510(k) number 2015-05-29 Class II Nonconforming Material/Component GE Healthcare has recently become aware of a potential safety issue that may result in the loss of C-arm gantry motion capability on fluoroscopic Interventional imaging systems (Innova, Optima and Discovery). 3 2015-10-28
GE Medical Systems, LLC 510(k) number 2015-05-29 Class II Nonconforming Material/Component GE Healthcare has recently become aware of a potential safety issue that may result in the loss of C-arm gantry motion capability on fluoroscopic Interventional imaging systems (Innova, Optima and Discovery). 4 2015-10-28
GE Medical Systems, LLC 510(k) number 2015-05-29 Class II Nonconforming Material/Component GE Healthcare has recently become aware of a potential safety issue that may result in the loss of C-arm gantry motion capability on fluoroscopic Interventional imaging systems (Innova, Optima and Discovery). 13 2015-10-28
GE Medical Systems, LLC 510(k) number 2015-05-29 Class II Nonconforming Material/Component GE Healthcare has recently become aware of a potential safety issue that may result in the loss of C-arm gantry motion capability on fluoroscopic Interventional imaging systems (Innova, Optima and Discovery). 1 2015-10-28
GE Medical Systems, LLC 510(k) number 2015-05-29 Class II Nonconforming Material/Component GE Healthcare has recently become aware of a potential safety issue that may result in the loss of C-arm gantry motion capability on fluoroscopic Interventional imaging systems (Innova, Optima and Discovery). 6 2015-10-28
GE Medical Systems, LLC 510(k) number 2015-05-29 Class II Nonconforming Material/Component GE Healthcare has recently become aware of a potential safety issue that may result in the loss of C-arm gantry motion capability on fluoroscopic Interventional imaging systems (Innova, Optima and Discovery). 2 2015-10-28
GE Medical Systems, LLC 510(k) number 2015-05-29 Class II Nonconforming Material/Component GE Healthcare has recently become aware of a potential safety issue that may result in the loss of C-arm gantry motion capability on fluoroscopic Interventional imaging systems (Innova, Optima and Discovery). 10 2015-10-28
GE Medical Systems, LLC 510(k) number 2015-05-29 Class II Nonconforming Material/Component GE Healthcare has recently become aware of a potential safety issue that may result in the loss of C-arm gantry motion capability on fluoroscopic Interventional imaging systems (Innova, Optima and Discovery). 17 2015-10-28
GE Medical Systems, LLC 510(k) number 2015-05-29 Class II Nonconforming Material/Component GE Healthcare has recently become aware of a potential safety issue that may result in the loss of C-arm gantry motion capability on fluoroscopic Interventional imaging systems (Innova, Optima and Discovery). 6 2015-10-28
GE Medical Systems, LLC 510(k) number 2015-05-29 Class II Nonconforming Material/Component GE Healthcare has recently become aware of a potential safety issue that may result in the loss of C-arm gantry motion capability on fluoroscopic Interventional imaging systems (Innova, Optima and Discovery). 3 2015-10-28
GE Medical Systems, LLC 510(k) number 2015-05-29 Class II Nonconforming Material/Component GE Healthcare has recently become aware of a potential safety issue that may result in the loss of C-arm gantry motion capability on fluoroscopic Interventional imaging systems (Innova, Optima and Discovery). 8 2015-10-28
GE Medical Systems, LLC 510(k) number 2015-05-29 Class II Nonconforming Material/Component GE Healthcare has recently become aware of a potential safety issue that may result in the loss of C-arm gantry motion capability on fluoroscopic Interventional imaging systems (Innova, Optima and Discovery). 7 2015-10-28
GE Medical Systems, LLC 510(k) number 2014-09-04 Class II Device Design Loss of imaging – affected devices may experience a loss of real-time interventional imaging and/or gantry motion due to the Electromagnetic (EM) sensitivity of the grid detection mechanism. 10 (4 US, 6 OUS). 2015-01-21
GE Medical Systems, LLC 510(k) number 2013-11-15 Class II Software design GE Healthcare has recently become aware of a potential safety issue with respect to IGS 630 Imaging Systems.
While performing fluoroscopy on a biplane fluoroscopy unit, there is a potential for loss of the x-ray imaging function when the us
28 (12 US, 16 OUS). 2014-03-28

Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.