Veradius

Veradius is identified as an interventional fluoroscopic X-ray system. The filing describes a counterbalanced C-arm with a flat-detector X-ray system and a mobile viewing station. Its…

Registry build 21.09.26

Veradius is identified as an interventional fluoroscopic X-ray system. The filing describes a counterbalanced C-arm with a flat-detector X-ray system and a mobile viewing station.

Its stated use is radiological guidance and visualization during diagnostic, interventional and surgical procedures by trained healthcare professionals, excluding neonates.

Listed Item

Mobile, diagnostic X-ray imaging and viewing system with Flat Detector. Used for radiological guidance and visualization during diagnostic, interventional, and surgical procedures.

Clinical Setting and Applications

The filing says the system may be used inside or outside the operating room, in sterile or non-sterile healthcare settings. Its listed application areas are orthopedic, neuro, abdominal, vascular, thoracic and cardiac, within device limits.

5manufacturers in China registered for this device type
2catalogue numbers in this listing
29510(k) clearances naming this product code
40recalls on this code

Who Else Makes It

Concentrated in Beijing (2), Shanghai (2), Guangdong (1).

Chinese manufacturers registered for FDA product code OWB
Manufacturer City Province Registered as Device types
GE Hangwei Medical Systems Co., Ltd. Beijing Beijing Specification developer 22
GE Healthcare (China) Co., Ltd. Beijing Beijing Specification developer 22
Shanghai United Imaging Healthcare Co., Ltd. Shanghai Shanghai Manufacturer 12
Siemens Shanghai, Medical Equipment Ltd. Shanghai Shanghai Contract manufacturer 10
Siemens Shenzhen Magnetic Resonance Ltd. Shenzhen Guangdong Contract manufacturer 8

Record

Regulatory record for Veradius
Listed by Philips Medical Systems Nederland B.V., Netherlands
FDA product code OWB
Device type Interventional Fluoroscopic X-Ray System
Device class Class 2
Regulation 21 CFR 892.1650
Establishment Contract manufacturer, Remanufacturer, Specification developer, Complaint-file establishment, Manufacturer, Exporter
510(k) K133819
Record published 2016-09-24
Record revised 2025-12-29 (version 7)

Specifications

Specifications
MRI safety Labeling does not contain MRI Safety Information
Prescription Prescription only
Label carries serial number

Catalogue Numbers

Catalogue numbers for Veradius
Catalogue number Size Status
Veradius Unity In distribution
Veradius In distribution

Same Product Code

Other products under the same FDA product code
Product Sold by Catalogue numbers
TrackX Trackx Technology Inc. 86
Trinias Shimadzu Corp. 54
Allura Xper FD Philips Medical Systems Nederland B.V. 21
Azurion Philips Medical Systems Nederland B.V. 19
LessRay Nuvasive, Inc. 16
Intelli-C Accessory Nrt – Nordisk Røntgen Teknik A/S 13
Pulse Nuvasive, Inc. 12
O-ARM® Medtronic Navigation, Inc. 11
Genoray Genoray Co., Ltd. 10
Fujifilm Healthcare Americas Corporation A.T.S. Applicazione Tecnologie Speciali SRL 7
Fujifilm Medical Systems U.S.A. Inc. A.T.S. Applicazione Tecnologie Speciali SRL 6
Stephanix A.T.S. Applicazione Tecnologie Speciali SRL 6

Showing 12 of 122. All products under Interventional Fluoroscopic X-Ray System

510(K) Clearances

FDA 510(k) clearances naming this product code
Clearance Decision date
K254205 2026-08-07
K232420 2024-02-16
K212813 2021-10-01
K201583 2021-02-09
K200917 2020-05-01
K183101 2018-12-31
K183040 2018-12-06
K181830 2018-08-02
K172822 2017-11-22
K172307 2017-10-24
K170130 2017-06-02
K163715 2017-01-26
K162859 2016-12-02
K160455 2016-08-25
K160147 2016-08-10
K161563 2016-07-29
K161839 2016-07-29
K151598 2015-08-17
K142273 2015-01-30
K142126 2014-12-09
K133819 2014-07-10
K140138 2014-06-10
K133292 2014-03-05
K132147 2013-12-16
K132305 2013-10-30
K130893 2013-09-06
K131263 2013-07-24
K130638 2013-06-28
K121772 2013-03-21

Recalls Naming This Product

2 recalls name this product by its own identifiers, the UDI on its label.

Recalls naming the identifiers of Veradius
Recalled by Identified by Initiated FDA class Root cause Reason Quantity
Philips North America LLC UDI-DI 2025-02-25 Class II Labeling design A wireless foot switch pedal may get stuck in the active position when the user releases the pedal, resulting in the emission of unintended radiation.
Philips North America LLC UDI-DI 2021-11-12 Class II Device Design Instructions for Use do not specify the maximum surface temperature of the X-ray tank (assembly housing the X-ray tube, beam filter & cooling oil) and the image intensifier/detector of the system as required an result in burns 462 units

Recalls

40 of these are Class II. Class II means a temporary or reversible health consequence. Volume alone is not a quality signal: a manufacturer that ships more recalls more.

Recalls referencing this product code
Initiated FDA class Root cause Reason Quantity
2026-07-16 Class II Software design Two software related issues affecting Philips Azurion R3.0 and R3.1 systems:
1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate treatment
and
2) Potential loss of geo
2016 units
2026-07-16 Class II Software design Two software related issues affecting Azurion R2.2.10 Systems:
1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate treatment and
2) Potential loss of geometry movements
10872 units
2026-06-29 Class II Software design Firmware issue potentially leading to loss of geometry movements. 4,252 units
2026-06-17 Class II Under Investigation by firm Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives. 3,341 units
2026-06-17 Class II Under Investigation by firm Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives. 5,012 units
2026-06-17 Class II Under Investigation by firm Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives. 257 units
2026-05-18 Class II Radiation Control for Health and Safety Act Philips Allura Xper and MultiDiagnost-Eleva systems
Hand switch button not fully releasing potentially leading to loss of imaging functionality or unintended
radiation exposure and additional contrast injection
2026-04-30 Class II Under Investigation by firm It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in
performance as they age, particularly beyond six years of service. Issues with an HDD may, depending
on the specific system PC in which the aff
2026-04-30 Class II Under Investigation by firm It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in
performance as they age, particularly beyond six years of service. Issues with an HDD may, depending
on the specific system PC in which the aff
3069 units (872 US, 2224 OUS)
2026-04-28 Class II Component design/selection The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage.
Oil leakage may affect the cooling performance of the X-ray tube. Whe
3314 units (816 US, 2498 OUS)
2026-04-28 Class II Component design/selection The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage.
Oil leakage may affect the cooling performance of the X-ray tube. Whe
25 units (11 US, 14 OUS)
2026-04-28 Class II Component design/selection The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage.
Oil leakage may affect the cooling performance of the X-ray tube. Whe
1651 units (592 US, 1359 OUS)
2026-04-28 Class II Device Design Potential safety issue with Philips Azurion systems, where longitudinal and transverse table movements may be impacted due to mechanical wear of the Float Tabletop control on the control module. 5,537 units
2026-03-31 Class II Device Design Potential for the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may loosen and/or break over time due to the forces applied when the monitor is moved or rotated. 1,718 units
2026-03-03 Class II Device Design Philips has identified instances where X-ray imaging may not be initiated or may occur
intermittently when using the wired foot switch.
296 units (68 US, 228 OUS)
2026-03-03 Class II Device Design Philips has identified instances where X-ray imaging may not be initiated or may occur
intermittently when using the wired foot switch.
5 units (OUS only)
2026-03-03 Class II Device Design Philips has identified instances where X-ray imaging may not be initiated or may occur
intermittently when using the wired foot switch.
1523 units (322 US, 1201 OUS)
2026-03-03 Class II Device Design Philips has identified instances where X-ray imaging may not be initiated or may occur
intermittently when using the wired foot switch.
461 units (198 US, 263 OUS)
2026-03-03 Class II Device Design Philips has identified instances where X-ray imaging may not be initiated or may occur
intermittently when using the wired foot switch.
345 units (10 US, 335 OUS)
2026-03-03 Class II Device Design Philips has identified instances where X-ray imaging may not be initiated or may occur
intermittently when using the wired foot switch.
111 units (31 US, 80 OUS)
2026-03-03 Class II Device Design Philips has identified instances where X-ray imaging may not be initiated or may occur
intermittently when using the wired foot switch.
5084 units (1616 US, 3468 OUS)
2026-03-03 Class II Device Design Philips has identified instances where X-ray imaging may not be initiated or may occur
intermittently when using the wired foot switch.
85 units (20 US, 65 OUS)
2026-03-03 Class II Device Design Philips has identified instances where X-ray imaging may not be initiated or may occur
intermittently when using the wired foot switch.
2 units (2 US, 6 OUS)
2026-03-03 Class II Device Design Philips has identified instances where X-ray imaging may not be initiated or may occur
intermittently when using the wired foot switch.
2291 units (493 US, 1798 OUS)
2026-03-03 Class II Device Design Philips has identified instances where X-ray imaging may not be initiated or may occur
intermittently when using the wired foot switch.
10 units (8 US, 2 OUS)
2026-03-03 Class II Device Design Philips has identified instances where X-ray imaging may not be initiated or may occur
intermittently when using the wired foot switch.
920 units (168 US, 752 OUS)
2026-03-03 Class II Device Design Philips has identified instances where X-ray imaging may not be initiated or may occur
intermittently when using the wired foot switch.
196 units (73 US, 223 OUS)
2026-03-03 Class II Device Design Philips has identified instances where X-ray imaging may not be initiated or may occur
intermittently when using the wired foot switch.
4 units (2 US, 2 OUS)
2026-03-03 Class II Device Design Philips has identified instances where X-ray imaging may not be initiated or may occur
intermittently when using the wired foot switch.
5167 units (1728 US, 3439 OUS)
2026-03-03 Class II Device Design Philips has identified instances where X-ray imaging may not be initiated or may occur
intermittently when using the wired foot switch.
670 units (106 US, 564 OUS)
2026-03-03 Class II Device Design Philips has identified instances where X-ray imaging may not be initiated or may occur
intermittently when using the wired foot switch.
798 units (42 US, 756 OUS)
2026-03-03 Class II Device Design Philips has identified instances where X-ray imaging may not be initiated or may occur
intermittently when using the wired foot switch.
298 units (27 US, 271 OUS)
2026-03-03 Class II Device Design Philips has identified instances where X-ray imaging may not be initiated or may occur
intermittently when using the wired foot switch.
1994 units (360 US, 1634 OUS)
2026-03-03 Class II Device Design Philips has identified instances where X-ray imaging may not be initiated or may occur
intermittently when using the wired foot switch.
2 units (1 US, 1 OUS)
2026-03-03 Class II Device Design Philips has identified instances where X-ray imaging may not be initiated or may occur
intermittently when using the wired foot switch.
15 units (7 US, 8 OUS)
2026-03-03 Class II Device Design Philips has identified instances where X-ray imaging may not be initiated or may occur
intermittently when using the wired foot switch.
112 units (138 US, 422 OUS)
2026-01-26 Class II Device Design Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed – even when a table lock is active. 291 units (27 US, 264 OUS)
2026-01-26 Class II Device Design Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed – even when a table lock is active. 802 units (45 US, 757 OUS)
2026-01-26 Class II Device Design Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed – even when a table lock is active. 655 units (106 US, 549 OUS)
2026-01-26 Class II Device Design Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed – even when a table lock is active. 1503 units (326 US, 1177 OUS)

Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.

Where China Actually Supplies

Every device page names the company that lists it and the factories registered to make the same thing. These are the sections with the most of them.

5,100manufacturers registered in China
2,581device types they are registered for
340,248device records, each naming its maker
31provinces with a registered manufacturer