Visao®
Drill 3334800 Visao 80K Handpiece Rohs
Who Else Makes It
Concentrated in Shanghai (1), Guangdong (1).
| Manufacturer | City | Province | Registered as | Device types |
|---|---|---|---|---|
| Jabil Circuit (Shanghai) Ltd. | Shanghai | Shanghai | Contract manufacturer | 31 |
| Guangdong Jinme Medical Technology Co., Ltd. | Foshan | Guangdong | Manufacturer | 13 |
Record
| Listed by | Medtronic Xomed, Inc., United States |
|---|---|
| FDA product code | ERL |
| Device type | Drill, Surgical, Ent (Electric Or Pneumatic) Including Handpiece |
| Device class | Class 2 |
| Regulation | 21 CFR 874.4250 |
| Establishment | Specification developer, Repacker / relabeller, Complaint-file establishment, Manufacturer, Exporter |
| 510(k) | K011321 |
| Record published | 2016-04-30 |
| Record revised | 2022-09-16 (version 9) |
Specifications
| MRI safety | Labeling does not contain MRI Safety Information |
|---|---|
| Sterilisation | Ethylene Oxide, Moist Heat or Steam Sterilization |
| Supplied sterile | Yes |
| Single use | Yes |
| Prescription | Prescription only |
| Label carries | lot or batch number, serial number |
Catalogue Numbers
| Catalogue number | Size | Per pack | Status |
|---|---|---|---|
| 3334800 | In distribution | ||
| 3334610 | In distribution | ||
| 3334895 | In distribution | ||
| 3334800LO | In distribution | ||
| 3334800 | In distribution | ||
| 3334635 | In distribution | ||
| 3334800T | In distribution | ||
| 3334800 | In distribution | ||
| 3334895 | In distribution | ||
| 3334800LO | In distribution | ||
| 3334625 | In distribution | ||
| 3334635 | In distribution | ||
| 3334610 | In distribution | ||
| 3334625 | In distribution | ||
| 3334800T | In distribution | ||
| 3318606 | 2 | In distribution | |
| 3334800T | In distribution |
Same Product Code
| Product | Sold by | Catalogue numbers |
|---|---|---|
| ENT MR8 | Medtronic Powered Surgical Solutions | 234 |
| Drill, Surgical, Ent (Electric Or Pneumatic) Including Handpiece | Anspach Effort, The | 214 |
| Drill, Surgical, Ent (Electric Or Pneumatic) Including Handpiece | Karl Storz GmbH & Co. KG | 208 |
| Drill, Surgical, Ent (Electric Or Pneumatic) Including Handpiece | Synthes Gmbh | 126 |
| Midas Rex | Medtronic Powered Surgical Solutions | 97 |
| Drill, Surgical, Ent (Electric Or Pneumatic) Including Handpiece | Stryker GmbH | 87 |
| Midas REX Microsaws | Medtronic Powered Surgical Solutions | 75 |
| Piezomed Pro Accessory | W&H Dentalwerk Bürmoos GmbH | 50 |
| Midas Rex Clearview™ | Medtronic Xomed, Inc. | 47 |
| Zethon | Zethon Ltd | 39 |
| NSK Sterile Burs/Drills | Nakanishi Inc. | 34 |
| Drill, Surgical, Ent (Electric Or Pneumatic) Including Handpiece | Medtronic Xomed, Inc. | 29 |
Showing 12 of 169. All products under Drill, Surgical, Ent (Electric Or Pneumatic) Including Handpiece
510(K) Clearances
| Clearance | Decision date |
|---|---|
| K150728 | 2015-08-14 |
| K081277 | 2008-09-05 |
| K073255 | 2008-03-24 |
| K041523 | 2004-07-23 |
| K041413 | 2004-07-12 |
| K011321 | 2001-06-26 |
| K002224 | 2000-12-04 |
| K963246 | 1996-09-23 |
Recalls Naming This Product
3 recalls name this product by its own identifiers, the 510(k) it is cleared under.
| Recalled by | Identified by | Initiated | FDA class | Root cause | Reason | Quantity | Closed |
|---|---|---|---|---|---|---|---|
| Medtronic Xomed, Inc. | 510(k) number | 2021-12-15 | Class II | Under Investigation by firm | As a result of Medtronic’s integrated power console (authentication number: 222ACBZX00018000) being used with software version: v2.7.3.0, it was discovered that there was a defect that resulted in a difference between the set number of rota | 1447 units | 2023-06-09 |
| Medtronic Xomed, Inc. | 510(k) number | 2021-08-31 | Class II | Process control | There was an increase in complaints related to tip breaks and wobble/vibration. | 274,991 total devices | |
| Medtronic Powered Surgical Solutions | 510(k) number | 2016-05-13 | Class II | Device Design | Medtronic Powered Surgical Solutions testing of the Sagittal Saw Rocker Arm Seal identified a potential for leakage of fluid into the motor during use. | 368 devices | 2017-06-13 |
Recalls
1 of these is Class II. Class II means a temporary or reversible health consequence. Volume alone is not a quality signal: a manufacturer that ships more recalls more.
| Initiated | FDA class | Root cause | Reason | Quantity | Closed |
|---|---|---|---|---|---|
| 2020-06-12 | Class II | Software design | During internal testing execution of the next generation of Integrated Power Console (IPC) prototype it was noted that the M5 Microdebrider was rotating at a higher speed than the set value in the cosole. | 673 units | 2023-04-20 |
| 2009-07-17 | Not classified | Software design | An error code (Code 13) is displayed on some of the consoles when a Medtronic Stylus or EHS handpiece is connected to the IPC indicating that the handpiece has not been recognized by the console. | 493 | 2010-01-20 |
Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.