| Siemens Medical Solutions USA, Inc. | 510(k) number | 2018-05-25 | Class II | Process control | Due to a defective sealing, coolant may inflow into the electrical parts of the system cabinet, which could cause a short circuit and cause the system to shut down suddenly during an ongoing procedure | 3 US | 2019-05-16 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2018-05-01 | Class II | Device Design | After the Large Display returns from power save mode, it may not show an image, and stay black without showing an error message although X-ray is still possible. The problem does not occur during an ongoing procedure. If the problem occurs, | 68 | 2019-05-16 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2018-05-01 | Class II | Device Design | After the Large Display returns from power save mode, it may not show an image, and stay black without showing an error message although X-ray is still possible. The problem does not occur during an ongoing procedure. If the problem occurs, | 2 | 2019-05-16 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2018-05-01 | Class II | Device Design | After the Large Display returns from power save mode, it may not show an image, and stay black without showing an error message although X-ray is still possible. The problem does not occur during an ongoing procedure. If the problem occurs, | 73 | 2019-05-16 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2018-05-01 | Class II | Device Design | After the Large Display returns from power save mode, it may not show an image, and stay black without showing an error message although X-ray is still possible. The problem does not occur during an ongoing procedure. If the problem occurs, | 14 | 2019-05-16 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2018-05-01 | Class II | Device Design | After the Large Display returns from power save mode, it may not show an image, and stay black without showing an error message although X-ray is still possible. The problem does not occur during an ongoing procedure. If the problem occurs, | 1 | 2019-05-16 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2018-05-01 | Class II | Device Design | After the Large Display returns from power save mode, it may not show an image, and stay black without showing an error message although X-ray is still possible. The problem does not occur during an ongoing procedure. If the problem occurs, | 1 | 2019-05-16 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2018-05-01 | Class II | Device Design | After the Large Display returns from power save mode, it may not show an image, and stay black without showing an error message although X-ray is still possible. The problem does not occur during an ongoing procedure. If the problem occurs, | 56 | 2019-05-16 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2017-09-28 | Class II | Device Design | The possibility exists for the wireless foot switch to fail due to impacts by external factors, such as electrostatic discharge that exceed a certain intensity. If the foot switch fails, it will no longer be able to be used to release radia | 510 | 2020-03-09 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-12-08 | Class II | Software in the Use Environment | Siemens initiated a corrective action to address two possible, mutually independent causes of a system defect related to the following: – In Artis Systems with A100 Plus or A100G generators, an attempt to resume operation following detectio | 1,500 distributed Worldwide | 2019-04-09 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2016-11-01 | Class II | Software in the Use Environment | The possibility exists that the semi-automatic focus switch may not work as intended when pressing the footswitch multiple times. | 4,657 systems Worldwide | 2017-06-21 |
| Siemens Medical Solutions USA, Inc. | 510(k) number | 2015-05-22 | Class II | Software design | There exists a possible position sensor fault in the swivel base axis not being detected by the system software. When moving, the system could potentially exceed its usual speed, resulting in a collision. No injuries reported. | 475 | 2016-12-13 |