Incisive CT

Incisive CT is a whole-body computed tomography system that reconstructs X-ray transmission measurements into cross-sectional images. A continuously rotating tube and detector array acquire…

Registry build 21.09.26

Incisive CT is a whole-body computed tomography system that reconstructs X-ray transmission measurements into cross-sectional images. A continuously rotating tube and detector array acquire data from different angles as the patient couch moves through the gantry.

The system has a 72 cm bore and a 50 cm scan field of view. The console controls acquisition and supports image reconstruction, display, post-processing, storage and review.

Listed Item

The Incisive CT is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Clinical Applications

It is indicated for head, whole body, cardiac and vascular CT in patients of all ages. Low-dose lung cancer screening is supported only within established eligibility criteria of published screening programs or protocols.

8manufacturers in China registered for this device type
2catalogue numbers in this listing
8510(k) clearances naming this product code
4recalls on this code

Who Else Makes It

Concentrated in Beijing (2), Shanghai (2), Jiangsu (2). 2 of them are registered as contract manufacturers, able to build to a buyer’s own specification.

Chinese manufacturers registered for FDA product code JAK
Manufacturer City Province Registered as Device types
Philips Healthcare (Suzhou) Co., Ltd. Suzhou Jiangsu Contract manufacturer 5
Qisda Electronics (Suzhou) Co., Ltd. Suzhou Jiangsu Contract manufacturer 5
GE Hangwei Medical Systems Co., Ltd. Beijing Beijing Specification developer 22
GE Healthcare (China) Co., Ltd. Beijing Beijing Specification developer 22
Shanghai United Imaging Healthcare Co., Ltd. Shanghai Shanghai Manufacturer 12
Siemens Shanghai, Medical Equipment Ltd. Shanghai Shanghai Contract manufacturer 10
Neusoft Medical Systems Co., Ltd. Shenyang Liaoning Manufacturer 1
Philips Health Technology (China) Co., Ltd. Shenyang Branch Shenyang Liaoning Specification developer 1

Record

Regulatory record for Incisive CT
Listed by Philips Healthcare (Suzhou) Co., Ltd., China
FDA product code JAK
Device type System, X-Ray, Tomography, Computed
Device class Class 2
Regulation 21 CFR 892.1750
Establishment Contract manufacturer, Specification developer, Complaint-file establishment, Manufacturer
510(k) K180015
Record published 2021-12-30
Record revised 2025-07-04 (version 5)

Specifications

Specifications
MRI safety Labeling does not contain MRI Safety Information
Prescription Prescription only
Label carries serial number
Storage Handling environment temperature -20 to 50 °C; Handling environment humidity 20 to 80 % RH; Handling environment atmospheric pressure 50 to 106 KiloPascal; Storage environment temperature -15 to 45 °C; Storage environment humidity 20 to 80 % RH; Storage environment atmospheric pressure 50 to 106 KiloPascal
Distribution end date 2024-12-31

Catalogue Numbers

Catalogue numbers for Incisive CT
Catalogue number Size Status
Incisive CT In distribution
Incisive CT Withdrawn

Same Product Code

Other products under the same FDA product code
Product Sold by Catalogue numbers
Visicoil IZI Medical Products LLC 48
Visicoil MR IZI Medical Products LLC 41
Illumisite Covidien LP 25
Imbio LDA Imbio, LLC 24
Edge Covidien LP 20
Revolution GE Hangwei Medical Systems Co., Ltd. 20
Gold Fiducial IZI Medical Products LLC 20
superDimension Covidien LP 17
Lodox Lodox Systems (PTY) Ltd. 12
Revolution GE Medical Systems, L.L.C. 12
Toshiba Scanner Canon Inc. 10
syngo.CT Dual Energy Siemens Healthcare Limited 7

Showing 12 of 495. All products under System, X-Ray, Tomography, Computed

510(K) Clearances

FDA 510(k) clearances naming this product code
Clearance Decision date
K252992 2026-03-23
K232491 2024-05-03
K223311 2022-12-22
K212441 2022-04-27
K211168 2021-11-22
K203514 2021-06-17
K191136 2019-07-29
K180015 2018-03-20

Recalls Naming This Product

24 recalls name this product by its own identifiers, the UDI on its label or the 510(k) it is cleared under.

Recalls naming the identifiers of Incisive CT
Recalled by Identified by Initiated FDA class Root cause Reason Quantity Closed
Philips Healthcare (Suzhou) Co., Ltd. 510(k) number 2025-12-03 Class II Radiation Control for Health and Safety Act Potential for incomplete scan due to unstable connection inside of floating sensor. 35
Philips North America LLC 510(k) number 2025-10-29 Class II Under Investigation by firm If the screws of tube heat exchanger on the rotating scanner are not tightened well after replacing. This component may become detached and make contact with other components located within the Incisive CT systems during rotation. Other com 2341 units
Philips Medical Systems Technologies Ltd 510(k) number 2025-09-25 Class II Under Investigation by firm Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop.
Issue 2. When performing a helical/Axial scan with ORI (Dose Right index)/ DOM (Dose Modulation), t
105 units
Philips Medical Systems Technologies Ltd 510(k) number 2025-09-25 Class II Under Investigation by firm Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop.
Issue 2. When performing a helical/Axial scan with ORI (Dose Right index)/ DOM (Dose Modulation), t
165 units
Philips Medical Systems Technologies Ltd 510(k) number 2025-09-25 Class II Under Investigation by firm Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop.
Issue 2. When performing a helical/Axial scan with ORI (Dose Right index)/ DOM (Dose Modulation), t
1 unit (OUS only)
Philips North America LLC 510(k) number 2024-04-15 Class II Under Investigation by firm Metal mounting box on the rotating scanner on rotor (heat change box) located within the Incisive CT system, may become compromised, resulting in potential contact with other gantry components located within the Incisive CT system during ro 18 units
Philips North America LLC UDI-DI 2024-04-15 Class II Under Investigation by firm Metal mounting box on the rotating scanner on rotor (heat change box) located within the Incisive CT system, may become compromised, resulting in potential contact with other gantry components located within the Incisive CT system during ro 155 units
Philips North America LLC 510(k) number 2023-12-07 Class II Component design/selection If the Patient Interface Monitor (PIM) Cable is plugged/unplugged frequently as part of daily operation, the connector and wire may disconnect from the welding point, which may result in intermittent loss of the ECG signal. Loss of ECG sign 55 US; 1029 ROW
Philips Healthcare Group 510(k) number 2023-08-02 Class II Radiation Control for Health and Safety Act Philips has identified three software issues with compliance concerns to Subchapter J: Issue 1: Coronary Computed Tomography Angiography (CTA) scan may not be triggered at the optimal time. Issue 2: Use of aborted surview for scan planning 140 systems
Philips North America LLC 510(k) number 2023-04-25 Class II Under Investigation by firm Metal mounting box on the rotating scanner on rotor
(heat exchanger box) located within the Incisive CT system may become expelled and make contact with other components and cause injury to user or if detached a loud noise will be emitted a
36 US; 610 OUS
Philips North America LLC 510(k) number 2023-04-25 Class II Under Investigation by firm Metal mounting box on the rotating scanner on rotor
(heat exchanger box) located within the Incisive CT system may become expelled and make contact with other components and cause injury to user or if detached a loud noise will be emitted a
20 units OUS
Philips North America LLC 510(k) number 2023-04-25 Class II Under Investigation by firm Metal mounting box on the rotating scanner on rotor
(heat exchanger box) located within the Incisive CT system may become expelled and make contact with other components and cause injury to user or if detached a loud noise will be emitted a
53 units OUS
Philips North America LLC 510(k) number 2023-04-25 Class II Under Investigation by firm Metal mounting box on the rotating scanner on rotor
(heat exchanger box) located within the Incisive CT system may become expelled and make contact with other components and cause injury to user or if detached a loud noise will be emitted a
6 units OUS
Philips Headquarters Cambridge 510(k) number 2023-04-12 Class II Software design Potential for Incorrect Image Orientation resulting images may be flipped or reversed result in misdiagnosis, incorrect treatment of a condition, or additional radiation exposure if a rescan is required 7 units US and 76 unit OUS
Philips Headquarters Cambridge 510(k) number 2023-04-12 Class II Software design Potential for Incorrect Image Orientation resulting images may be flipped or reversed result in misdiagnosis, incorrect treatment of a condition, or additional radiation exposure if a rescan is required 22 units
Philips Headquarters Cambridge 510(k) number 2023-04-12 Class II Software design Potential for Incorrect Image Orientation resulting images may be flipped or reversed result in misdiagnosis, incorrect treatment of a condition, or additional radiation exposure if a rescan is required 1 unit
Philips Headquarters Cambridge 510(k) number 2023-04-12 Class II Software design Potential for Incorrect Image Orientation resulting images may be flipped or reversed result in misdiagnosis, incorrect treatment of a condition, or additional radiation exposure if a rescan is required 5 units
Philips North America LLC 510(k) number 2023-03-20 Class II Under Investigation by firm Multi-Function Foot Switch Unload Pedal Foot Entrapment-Pressing the Unload Pedal of the foot switch to perform the unload function may cause entrapment of the operator’s foot and injury 10 units US; 36 units OUS
Philips North America LLC 510(k) number 2023-03-20 Class II Under Investigation by firm Multi-Function Foot Switch Unload Pedal Foot Entrapment-Pressing the Unload Pedal of the foot switch to perform the unload function may cause entrapment of the operator’s foot and injury 3 units US ;2 units OUS
Philips Healthcare (Suzhou) Co., Ltd. 510(k) number 2022-12-23 Class II Software design Multiple software issues which have the potential to lead to misdiagnosis due to image artifacts or incorrect image orientation labels, or the need for a CT rescan.

These include: 1. Artifacts may occur on perfusion images, 2. Orientation

828
Philips North America LLC 510(k) number 2022-03-29 Class II Software design Three software issues affecting incorrect image display, error interpreting patient images due to image reporting, and patient exposure to incorrect image/function during clinical use. 330 units
Philips Healthcare (Suzhou) Co., Ltd. 510(k) number 2021-02-24 Class II Process control Philips has internally detected an issue with the actuators used in the Incisive CT couch, which could result in the couch failure to self-lock, and unexpectedly moving slowly downward to its limit position. 109 systems 2023-02-17
Philips North America LLC 510(k) number 2021-02-22 Class II Process control Actuators in the Incisive gantry could result in the gantry failure to self-lock if both of the 2 sides actuators fail.

In some cases in systems with affected components, the gantry may tilt forward slowly, in a worst case, reaching to the

105 units (1 US and 104 OUS) 2023-02-09
Philips North America LLC 510(k) number 2020-12-24 Class II Software design Patient tabletop moved out to the home position during scan initialization, may cause operator/bystander staying by the rear of the table to collide with the moved tabletop and be injured 428 units

Recalls

1 of these is Class II. Class II means a temporary or reversible health consequence. Volume alone is not a quality signal: a manufacturer that ships more recalls more.

Recalls referencing this product code
Initiated FDA class Root cause Reason
2026-06-26 Class II Radiation Control for Health and Safety Act Potential of failing to trigger clinical scan during Bolus Tracking.

Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.