Delivery Analysis

Software, Delivery Analysis, Version 1.2.1

Registry build 21.09.26
5manufacturers in China registered for this device type
1catalogue number
9510(k) clearances naming this product code
12recalls on this code

Who Else Makes It

Concentrated in Jiangsu (2), Fujian (1), Guangdong (1). One is registered as a contract manufacturer, able to build to a specification.

Chinese manufacturers registered for FDA product code MUJ
Manufacturer City Province Registered as Device types
Philips Healthcare (Suzhou) Co., Ltd. Suzhou Jiangsu Contract manufacturer 5
Manteia Technologies Co., Ltd. Xiamen Fujian Manufacturer 3
Masep Medical Science & Technology DEVELOPMENT(SHENZHEN)Co. Shenzhen Guangdong Manufacturer 2
Suzhou Linatech Medical Science AND Technology Co., Ltd. Suzhou Jiangsu Manufacturer 2
Varian Medical Systems China Co., Ltd. Beijing Beijing Manufacturer 2

Record

Regulatory record for Delivery Analysis
Listed by Accuray Incorporated, United States
FDA product code MUJ
Device type System, Planning, Radiation Therapy Treatment
Device class Class 2
Regulation 21 CFR 892.5050
Establishment Manufacturer
510(k) K143269
Record published 2020-02-11

Specifications

Specifications
MRI safety Labeling does not contain MRI Safety Information
Prescription Prescription only

Catalogue Numbers

Catalogue numbers for Delivery Analysis
Catalogue number Size Status
1062213 In distribution

Same Product Code

Other products under the same FDA product code
Product Sold by Catalogue numbers
RayStation RaySearch Laboratories AB (publ) 29
System, Planning, Radiation Therapy Treatment Nucletron B.V. 17
Panther TPS Prowess, Inc. 14
RayCare RaySearch Laboratories AB (publ) 12
Accuray Precision® Treatment Planning System Accuray Incorporated 7
Monaco Elekta Inc. 7
Eclipse Treatment Planning System for External Beam Radiation Therapy Varian Medical Systems, Inc 6
syngo.via RT Image Suite Siemens Healthcare Limited 5
Eclipse Treatment Planning System Varian Medical Systems, Inc 5
Corvus 09 Best Nomos 4
Monaco Elekta Solutions AB 4
XKnife® Integra Lifesciences Corporation 4

Showing 12 of 98. All products under System, Planning, Radiation Therapy Treatment

510(K) Clearances

FDA 510(k) clearances naming this product code
Clearance Decision date
K161136 2016-06-24
K152488 2015-09-28
K143269 2015-04-14
K091146 2009-05-01
K062967 2006-10-30
K062514 2006-09-28
K052325 2005-09-14
K051941 2005-08-09
K032345 2003-09-03

Recalls

The FDA has not published a severity grade for these. Volume alone is not a quality signal: a manufacturer that ships more recalls more.

Recalls referencing this product code
Initiated FDA class Root cause Reason Quantity Closed
2011-07-13 Not classified Software design As a result of an internal review, TomoTherapy has identified an issue with HiArt versions 4.0.x, and HD versions 1.0.x that we would like to bring to your attention.
During DICOM export of plan level images with a non square exported Fiel
346 2014-03-18
2011-07-07 Not classified Device Design TomoTherapy Inc. is sending this Field Safety Notice to make you aware of an anomaly which may affect the performance of the TomoMobile Hi-Art System.

TomoTherapy has discovered that when attempting to open the TomoMobile shielding doors,

4 2012-06-29
2011-01-26 Not classified Software design It was determined that the Treatment Planning Station (TPS) can potentially under dose. 326 2014-03-18
2010-11-08 Not classified Software design TomoTherapy Inc. is sending this Field Safety Notice to make you aware of an anomaly which may affect the performance of the Hi-Art System and make it so the user cannot generate a Completion Procedure. 40 (21US; 19 OUS) 2012-02-04
2010-07-06 Not classified Device Design Latches responsible for securing the cover to the device may come lose if not properly secured and the cover may unexpectedly drop off. A design change is planned. 207 units 2011-10-04
2010-06-23 Not classified Software design An issue has been identified, as a result of input received from one TomoTherapy Hi-Art Treatment System customer.

With a particular workflow, the Planning Station’s Draft and/or final Plan Report can provide incorrect dose statistics.

29 2011-12-17
2010-02-22 Not classified Software design In some cases, the patient’s diagnostic CT image is narrower than the Hi-Art radiotherapy couch image. 7 2011-02-10
2010-01-15 Not classified Software design When a user modified the align center of an existing plan during the treatment planning process the effective depths of each beam in the plan were not updated by the system. These depths were incorrect and led to an incorrect dose calculati 99 units 2011-02-11
2009-12-29 Not classified Employee error If electron density values are left empty, the calculation of radiation dose in a patient will be modeled as air-like density material rather than correct density. A plan may be created and saved, thus creating the risk of mistreatment. 71 units are subject to this action 2011-02-11
2009-12-01 Not classified Software design An issue was identified with the TomoTherapy HI-Art System.

In the event a patient or DQA plan has a moved image, roll adjustments applied during registration will be incorrect.

The Planning Station Plan Settings and DQA Setup tabs al

1 2011-12-17
2009-08-10 Not classified Software design TomoTherapy received one customer report indicating that after starting a patient procedure from the Status
Console, the radiation on light and the audible indicator on the Status Console came on as expected,
however the Operator Station (O
259 (169 US; 91 OUS) 2011-12-17
2007-03-21 Not classified Software design Software problem – Sagittal, oblique or coronal sliced MR Images imported from a non-Accuray manufactured device into the product may result in inverted image orientation. 25 units 2007-12-26

Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA.

Records are records: nothing here is an offer of sale.

Where China Actually Supplies

Every device page names the company that lists it and the factories registered to make the same thing. These are the sections with the most of them.

5,100manufacturers registered in China
2,581device types they are registered for
340,248device records, each naming its maker
31provinces with a registered manufacturer