| Biomet, Inc. | 510(k) number | 2024-03-13 | Class II | Process change control | Possible damage to the Vanguard Knee System AS Tibial Bearing caused during the manufacturing process. | 19 units (1 US, 18 OUS) | |
| Biomet, Inc. | 510(k) number | 2022-12-29 | Class II | Labeling Change Control | Components Incorrectly labeled as either smaller or larger and incorrect side (right/left). | 157 units | |
| Biomet, Inc. | 510(k) number | 2022-12-29 | Class II | Labeling Change Control | Components Incorrectly labeled as either smaller or larger and incorrect side (right/left). | 91 units | |
| Biomet, Inc. | 510(k) number | 2020-04-17 | Class II | Process control | Potential presence of elevated endotoxin levels that exceed the specification limit | 11 units | 2021-06-25 |
| Biomet, Inc. | 510(k) number | 2020-04-17 | Class II | Process control | Potential presence of elevated endotoxin levels that exceed the specification limit | 107 units | 2021-06-25 |
| Biomet, Inc. | 510(k) number | 2020-04-17 | Class II | Process control | Potential presence of elevated endotoxin levels that exceed the specification limit | 91 units | 2021-06-25 |
| Biomet, Inc. | 510(k) number | 2018-12-21 | Class II | Under Investigation by firm | Vanguard Knee System Series-A Standard Patella a potential commingle that could result in the size of the implant not matching the size on the label. | 45 units | 2020-05-13 |
| Biomet, Inc. | 510(k) number | 2018-12-21 | Class II | Under Investigation by firm | Vanguard Knee System Series-A Standard Patella a potential commingle that could result in the size of the implant not matching the size on the label. | 90 units | 2020-05-13 |
| Biomet, Inc. | 510(k) number | 2018-10-22 | Class II | Labeling mix-ups | Label on the outer carton or the patient labels does not match label on the inner sterile packaging | 8 units | 2019-12-11 |
| Biomet, Inc. | 510(k) number | 2018-10-22 | Class II | Labeling mix-ups | Label on the outer carton or the patient labels does not match label on the inner sterile packaging | 8 units | 2019-12-11 |
| Biomet, Inc. | 510(k) number | 2018-05-24 | Class II | Under Investigation by firm | Two lots of tibial bearings were commingled. There is a possibility that a 14mm tibial bearing is packaged in a box labeled as 12 mm tibial bearing and vice versa. This may lead to extension of surgery to find another implant. | 48 | 2020-04-07 |
| Biomet, Inc. | 510(k) number | 2018-05-18 | Class II | Under Investigation by firm | One lot of 75mm tibial tray is potentially etched and labeled as a 79 mm tibial tray. This may lead to delay in surgery, or immediate revision surgery to remove an incorrectly sized tibial component after it has been cemented. | 16 | 2020-03-31 |
| Biomet, Inc. | 510(k) number | 2018-05-11 | Class II | Under Investigation by firm | The label master file was errantly set up as a 63/37mm instead of a 63/67mm. There is no 63/37 size offered, and the product is laser marked with the correct size. | 169 | 2019-11-21 |
| Biomet, Inc. | 510(k) number | 2017-10-03 | Class II | Error in labeling | The affected products are labeled as a 62.5mm LEFT Femur; however, the contents are a Vanguard 70mm RIGHT Femur. | 8 | 2018-08-23 |
| Biomet, Inc. | 510(k) number | 2017-02-17 | Class II | Process design | Endotoxin levels higher than process maximum limits were discovered in the identified polyethylene components. | 28,253 in total | 2018-01-29 |
| Biomet, Inc. | 510(k) number | 2011-10-04 | Not classified | Error in labeling | Correcting language on product labeling which incorrectly states patella buttons are intended for use with the AGC knee system. | | 2013-04-08 |
| Biomet, Inc. | 510(k) number | 2011-10-04 | Not classified | Error in labeling | Correcting language on product labeling which incorrectly states patella buttons are intended for use with the AGC knee system. | | 2013-04-08 |
| Biomet, Inc. | 510(k) number | 2011-10-04 | Not classified | Error in labeling | Correcting language on product labeling which incorrectly states patella buttons are intended for use with the AGC knee system. | | 2013-04-08 |
| Biomet, Inc. | 510(k) number | 2011-10-04 | Not classified | Error in labeling | Correcting language on product labeling which incorrectly states patella buttons are intended for use with the AGC knee system. | | 2013-04-08 |
| Biomet, Inc. | 510(k) number | 2011-10-04 | Not classified | Error in labeling | Correcting language on product labeling which incorrectly states patella buttons are intended for use with the AGC knee system. | | 2013-04-08 |
| Biomet, Inc. | 510(k) number | 2011-10-04 | Not classified | Error in labeling | Correcting language on product labeling which incorrectly states patella buttons are intended for use with the AGC knee system. | 149 | 2013-04-08 |
| Biomet, Inc. | 510(k) number | 2011-10-04 | Not classified | Error in labeling | Correcting language on product labeling which incorrectly states patella buttons are intended for use with the AGC knee system. | | 2013-04-08 |
| Biomet, Inc. | 510(k) number | 2011-10-04 | Not classified | Error in labeling | Correcting language on product labeling which incorrectly states patella buttons are intended for use with the AGC knee system. | | 2013-04-08 |
| Biomet, Inc. | 510(k) number | 2011-10-04 | Not classified | Error in labeling | Correcting language on product labeling which incorrectly states patella buttons are intended for use with the AGC knee system. | | 2013-04-08 |
| Biomet, Inc. | 510(k) number | 2011-10-04 | Not classified | Error in labeling | Correcting language on product labeling which incorrectly states patella buttons are intended for use with the AGC knee system. | | 2013-04-08 |